G
M
ENERAL
1.
DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digits
DP63000A-I: Reflective LCD with full viewing angle
DP63000B-I: Transmissive LCD with selectable red or green LED backlight,
viewing angle optimized. Display color change capability with output state
when using an output module.
2. POWER: Input voltage range is +9 to +28 VDC with short circuit and input
polarity protection. Must use a DP6-MLPS1 or a Class 2 or SELV rated power
supply.
MODEL
DISPLAY COLOR
NO.
DP63000A-I
---
DP63000B-I
Red (max intensity)
DP63000B-I
Green (max intensity)
3. INPUT RANGES: Jumper Selectable
D.C. Currents: 200 µA, 2 mA, 20 mA, or 200 mA
4. SIGNAL INPUTS:
ACCURACY
INPUT
INPUT
@23 °C, less
RANGE
IMPEDANCE
than 85% RH
µ
200
A
1.111 K
0.1% of span
2 mA
0.1% of span
111
20 mA
0.1% of span
11
200 mA
0.1% of span
1
5. OVERRANGE RATINGS, PROTECTION & INDICATION:
9 to 28 VDC power circuit is not isolated from the signal circuit.
Input Overrange Indication: "
Input Underrange Indication: "
Display Overrange/Underrange Indication: "
6. DISPLAY RESPONSE TIME: 500 msec min.
7. NORMAL MODE REJECTION: 60 dB 50/60 Hz
8. USER INPUT (USR): Programmable input. Connect terminal to common
(USR COMM) to activate function. Internal 10KΩ pull-up resistor to +9 to
28 VDC.
Threshold Levels: V
= 1.0 V max; V
IL
Response Time: 5 msec typ.; 50 msec debounce (activation and release)
2
9. MEMORY: Nonvolatile E
PROM memory retains all programming
parameters and max/min values when power is removed.
10. ENVIRONMENTAL CONDITIONS:
Operating Temperature Range for DP63000A-I: -35 to 75°C
Operating Temperature Range for DP63000B-I depends on display color
and intensity level as per below:
INTENSITY LEVEL
Red Display
Green Display
Storage Temperature: -35 to 85°C
Operating and Storage Humidity: 0 to 85% max. relative humidity (non-
condensing)
P
N
ART
UMBER
DESCRIPTION
DC Current Meter with reflective display
DC Current Meter with backlight display
Single Relay Output Card
Dual Sinking Open Collector Output card
RS485 Serial Communications Card
RS232 Serial Communications Card
Micro-Line Power Supply, 85 to 250 VAC
RS232 Programming Cable (DB9-RJ11)
RS485 Programming Cable (DB9-RJ11)
PC Configuration Software for Windows 98, ME, 2000, XP
S
ETER
PECIFICATIONS
INPUT CURRENT
INPUT CURRENT
@ 9 VDC WITHOUT
@ 9 VDC WITH
DP6-RLY0
DP6-RLY0
10 mA
40 mA
85 mA
115 mA
95 mA
125 mA
MAX
INPUT
RESOLUTION
COEFFICIENT
SIGNAL
15 mA
10 nA
70 ppm / °C
50 mA
.1 µA
70 ppm / °C
150 mA
1 µA
70 ppm / °C
500 mA
10 µA
70 ppm / °C
".
OLOL
".
ULUL
.....
.....
"/"-
"
= 2.4 V min; V
= 28 VDC
IH
MAX
TEMPERATURE
1 & 2
-35 to 75°C
3
-35 to 70°C
4
-35 to 60°C
5
-35 to 50°C
1 & 2
-35 to 75°C
3
-35 to 65°C
4
-35 to 50°C
5
-35 to 35°C
I
NFORMATION
*
Vibration According to IEC 68-2-6: 5 to 500 Hz, in X, Y, Z direction for 1.5
hours, 5 g's.
Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions
Altitude: Up to 2000 meters
11. CONNECTIONS: Wire clamping screw terminals
Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 5 inch-lbs (0.565 N-m) max.
12. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 requirements for
indoor use. Installation Category I, Pollution Degree 2. High impact plastic
case with clear viewing window. Panel gasket and mounting clip included.
13. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File #E313607, UL61010A-1, CSA 22.2 No. 61010-1
Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.
UL Listed, File # E313547, UL508, CSA C22.2 No. 14-M95
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment
TEMP.
for measurement, control, and laboratory use, Part 1.
IP65 Enclosure rating (Face only), IEC 529
Type 4X Indoor Enclosure rating (Face only), UL50
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN 61326: Electrical Equipment for Measurement,
Control and Laboratory use.
Immunity to Industrial Locations:
Electrostatic discharge
Electromagnetic RF fields
Fast transients (burst)
Surge
RF conducted interference
Power frequency magnetic fields
Emissions:
Emissions
Notes:
1. Criterion A: Normal operation within specified limits.
14. WEIGHT: 3.2 oz (100 g)
PART NUMBER
DP63000A-I
DP63000B-I
DP6-RLY0
DP6-SNK0
DP6-COM1
DP6-COM2
DP6-MLPS1
DP6-232-CABLE
DP6-485-CABLE
DP6-SOFT
4
EN 61000-4-2
Criterion A
4 kV contact discharge
8 kV air discharge
EN 61000-4-3
Criterion A
10 V/m
EN 61000-4-4
Criterion A
2 kV power
1 kV signal
EN 61000-4-5
Criterion A
1 kV L-L,
2 kV L&N-E power
EN 61000-4-6
Criterion A
3 V/rms
EN 61000-4-8
Criterion A
30 A/m
EN 55011
Class A

Advertisement