Connecting Analog Current Input Signals

Method 1

© National Instruments Corporation
M Series devices use a calibration method that requires some codes
(typically about 5% of the codes) to lie outside of the specified range. This
calibration method improves absolute accuracy, but it increases the nominal
resolution of input ranges by about 5% over what the formulas shown above
would indicate.
Choose an input range that matches the expected input range of your signal.
A large input range can accommodate a large signal variation, but reduces
the measurement resolution. Choosing a smaller input range improves the
measurement resolution, but may result in the input signal going out of
range.
For more information on programming these settings, refer to the
NI-DAQmx Help or the LabVIEW 8.x Help.
Table 4-1 shows the input ranges and resolutions supported by the NI 6238
and NI 6239 devices.
Table 4-1. Input Ranges for NI 6238/6239
Input Range
20 mA
When making signal connections, caution must be taken with the voltage
level of the signal going into the device. There are two types of connections
that can be made.
Method 1, shown in Figure 4-2, connects AI + and AI – inputs at a voltage
level with respect to AI GND. Verify that the voltage levels on the AI + and
AI – side do not exceed the common-mode input range of ±10 V.
Common-mode input range is the voltage input range with respect to
AI GND (AI + versus AI GND, AI – versus AI GND, or AI + versus AI –).
(20 mA – (–20 mA))
= 610 nA
16
2
Nominal Resolution Assuming
4-3
Chapter 4
Analog Input
5% Over Range
640 nA
NI 6238/6239 User Manual