Pxi Clock And Trigger Signals - National Instruments NI 6238 User Manual

Daq m series isolated current input/c urrent output devices
Hide thumbs Also See for NI 6238:
Table of Contents

Advertisement

RTSI, PFI, or
PXI_STAR Terminal
Filter Clock
(40 MHz)
Filtered Input

PXI Clock and Trigger Signals

Note PXI clock and trigger signals are only available on PXI devices. Other devices use
RTSI.
© National Instruments Corporation
Clocks Needed
Filter Setting
to Pass Signal)
125 ns
6.425 µs
2.55 ms
Disabled
The filter setting for each input can be configured independently. On power
up, the filters are disabled. Figure 10-3 shows an example of a low-to-high
transition on an input that has its filter set to 125 ns (N = 5).
1
1
2
3
4
Enabling filters introduces jitter on the input signal. For the 125 ns and
6.425 µs filter settings, the jitter is up to 25 ns. On the 2.55 ms setting, the
jitter is up to 10.025 µs.
When a PFI input is routed directly to RTSI, or a RTSI input is routed
directly to PFI, the M Series device does not use the filtered version of the
input signal.
Refer to the KnowledgeBase document, Digital Filtering with M Series
and CompactDAQ, for more information about digital filters and counters.
To access this KnowledgeBase, go to
code
.
rddfms
Chapter 10
Table 10-2. Filters
N (Filter
Pulse Width
Guaranteed to
Pass Filter
5
125 ns
257
6.425 µs
~101,800
2.55 ms
1
2
3
4
5
Figure 10-3. Filter Example
ni.com/info
10-7
Digital Routing and Clock Generation
Pulse Width
Guaranteed to
Not Pass Filter
100 ns
6.400 µs
2.54 ms
Filtered input goes high
when terminal is sampled
high on five consecutive
filter clocks.
and enter the info
NI 6238/6239 User Manual