National Instruments NI 6738 Calibration Procedure
National Instruments NI 6738 Calibration Procedure

National Instruments NI 6738 Calibration Procedure

Analog output devices and modules with ni-daqmx

Advertisement

Quick Links

PXIe-6738

Advertisement

Table of Contents
loading

Summary of Contents for National Instruments NI 6738

  • Page 1 PXIe-6738...
  • Page 2: Table Of Contents

    Worldwide Support and Services ..................... 27 Software Calibrating NI 6738/6739 devices and modules requires the installation of NI-DAQmx on the calibration system. You must use NI-DAQmx 17.6 or later. Refer to the NI-DAQmx Readme to ensure support for the device you want to calibrate. You can download NI-DAQmx from .
  • Page 3: Documentation

    Documentation Consult the following documents for information about your NI 6738/6739, NI-DAQmx, and your application software. All documents are available on , and help files install with ni.com the software. The DAQ Getting Started Guide for your device or module These guides provide instructions for installing the NI 6738/6739 with NI-DAQmx.
  • Page 4: Test Equipment

    Use with NI PXIe-6738/6739 chassis modules. Shielded NI SHC68-68-A2 — Use with NI 6738/6739 DAQ cable modules. NI SH68-C68-S — Use when connecting signals using the NI 6738 Adapter or NI 6739 Adapter. NI 6738/6739 Calibration Procedure | © National Instruments | 3...
  • Page 5 (x2) (Unterminated), 2 m If you are using the NI 4070, ensure that it is set to the 10 V range. † Locally constructed fixture to facilitate calibration using Calibration Executive. 4 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 6: Switch Maintenance

    Verify that all connections to the device are secure. • Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets. NI 6738/6739 Calibration Procedure | © National Instruments | 5...
  • Page 7: Device Setup

    Keep connections to the device as short as possible. Long cables and wires can act as antennae, which could pick up extra noise that would affect measurements. • Verify that all connections to the NI 6738/6739 are secure. • Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
  • Page 8: Calibration Procedure

    In the following procedures, Dev_name use the device name as it appears in MAX. Complete the following steps to set up the NI 6738/6739. Install the application software (if applicable). Install the NI-DAQmx driver software.
  • Page 9: Self-Calibration

    Change the system so that the temperature is within ±10 °C of the temperature recorded during the last external calibration. • Change the test limits to include the additional error due to temperature drift. Refer to your device specifications document for more information. 8 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 10: Verification

    These tables show all acceptable settings for the device. Analog Output Verification The NI 6738 has 32 analog outputs, AO <0..31>. The NI 6739 has 64 analog outputs, AO <0..63>. Note Table 7,...
  • Page 11 Connector 0, AO GND 22/23 (pin 62) AO 23 Connector 0, AO 23 (pin 28) Connector 0, AO GND 22/23 (pin 62) AO 24 Connector 0, AO 24 (pin 29) Connector 0, AO GND 24/25 (pin 30) 10 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 12 AO 49 Connector 1, AO 49 (pin 57) Connector 1, AO GND 48/49 (pin 24) AO 50 Connector 1, AO 50 (pin 58) Connector 1, AO GND 50/51 (pin 59) NI 6738/6739 Calibration Procedure | © National Instruments | 11...
  • Page 13 Analog Output — — Banana + (Red) — — Banana - (Black) AO 0 + Connector 0, — AO 0 (pin 10) AO 0 - Connector 0, — AO GND 0/1 (pin 11) 12 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 14 AO 7 + Connector 0, — AO 7 (pin 15) AO 7 - Connector 0, — AO GND 6/7 (pin 49) AO 8 + Connector 0, — AO 8 (pin 16) NI 6738/6739 Calibration Procedure | © National Instruments | 13...
  • Page 15 AO 14 - Connector 0, — AO GND 14/15 (pin 55) c15+ AO 15 + Connector 0, — AO 15 (pin 22) c15- AO 15 - Connector 0, — AO GND 14/15 (pin 55) 14 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 16 Connector 0, — AO 22 (pin 61) c22- AO 22 - Connector 0, — AO GND 22/23 (pin 62) c23+ AO 23 + Connector 0, — AO 23 (pin 28) NI 6738/6739 Calibration Procedure | © National Instruments | 15...
  • Page 17 AO 29 - Connector 0, — AO GND 28/29 (pin 33) c30+ AO 30 + Connector 0, — AO 30 (pin 67) c30- AO 30 - Connector 0, — AO GND 30/31 (pin 68) 16 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 18 AO 37 + Connector 1, — AO 37 (pin 47) AO 37 - Connector 1, — AO GND 36/37 (pin 14) AO 38 + Connector 1, — AO 38 (pin 48) NI 6738/6739 Calibration Procedure | © National Instruments | 17...
  • Page 19 AO 44 - Connector 1, — AO GND 44/45 (pin 20) c13+ AO 45 + Connector 1, — AO 45 (pin 54) c13- AO 45 - Connector 1, — AO GND 44/45 (pin 20) 18 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 20 Connector 1, — AO 52 (pin 26) c20- AO 52 - Connector 1, — AO GND 52/53 (pin 27) c21+ AO 53 + Connector 1, — AO 53 (pin 60) NI 6738/6739 Calibration Procedure | © National Instruments | 19...
  • Page 21 AO 59 - Connector 1, — AO GND 58/59 (pin 65) c28+ AO 60 + Connector 1, — AO 60 (pin 32) c28- AO 60 - Connector 1, — AO GND 60/61 (pin 33) 20 | ni.com | NI 6738/6739 Calibration Procedure...
  • Page 22 Call with the following parameters: DAQmxCreateTask • taskName: MyAOVoltageTask • taskHandle: &taskHandle Call with the following parameters. DAQmxCreateAOVoltageChan • taskHandle: taskHandle • physicalChannel: Dev_name/ao0 • nameToAssignToChannel: AOVoltageChannel • minVal: -10.0 NI 6738/6739 Calibration Procedure | © National Instruments | 21...
  • Page 23: Counter Verification

    You have finished verifying the analog output levels on your device. Counter Verification NI 6738/6739 devices have only one timebase to verify, so only Counter 0 needs to be checked. It is not possible to adjust this timebase, so only verification can be performed.
  • Page 24 If the counter verification procedure determines that the device is outside of the limits, refer to the Worldwide Support and Services section for assistance in returning the device to NI. NI 6738/6739 Calibration Procedure | © National Instruments | 23...
  • Page 25: Adjustment

    Adjustment The following procedure adjusts the analog output of your NI 6738/6739. At the end of each calibration procedure, these new constants are stored in the external calibration area of the EEPROM. These values are password-protected, which prevents the accidental access or modification of any calibration constants adjusted by the metrology laboratory.
  • Page 26: Eeprom Update

    10. Disconnect the DMM from the device. EEPROM Update When an adjustment procedure is completed, the NI 6738/6739 internal calibration memory (EEPROM) is immediately updated. If you do not want to perform an adjustment, you can update the calibration date without making...
  • Page 27: Test Limits

    The tables in this section list the specifications for PCIe/PXIe-6738 and PXIe-6739. The specifications are divided into analog output and counter/timer tables of values. Refer to the tables and specifications document for your device. Table 7. NI 6738/6739 Analog Output Values Range (V)
  • Page 28: Worldwide Support And Services

    For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and...