DISPlay <n>
Description Enables or disables the Meter's display. This setting is stored in volatile
memory.
Parameters <n>
Examples
DISP ON
Query
DISP?
DISPlay:TEXT(?) <quoted string >
Description Displays a message on the Meter's display. This setting is stored in volatile
memory.
The Meter must be remote before executing this command.
Parameters <quoted string> =
Examples
DISP:TEXT "Hello"
Query
DISP:TEXT?
DISPlay:TEXT:CLEar
Description Clears the displayed message on the Meter's display.
Examples
DISP:TEXT:CLE
SYSTem:BEEPer
Description Causes the Meter to beep once.
Examples
SYST:BEEP
SYSTem:BEEPer:STATe? <n>
Description Enables or disables the Meter's beeper. This setting is stored in volatile
memory. Disabling the beeper causes the Meter to not beep for the
following conditions:
Parameters <n>
Examples
SYST:BEEP:STAT ON Enables the beeper.
Query
SYST:BEEP:STAT?
SYSTem:DATE(?) <date>
Description Sets the system clock date.
= ON or 1
OFF or 0
Enables the Meter's display.
Retrieves the state of the display ( 0 = OFF and 1
= ON).
Note
Up to 12 characters. Additional characters are
truncated.
Hello is displayed in the Meter's display.
Retrieves the text sent to the Meter in a
DISP:TEXT command.
Meter's display message is cleared.
Meter beeps.
a new minimum or maximum is detected during statistics
a limit is exceeded during a limits test
a forward-biased diode is measured during a diode test
= ON or 1
OFF or 0
Retrieves the state of the Meter's beeper ( 0 =
OFF and 1 = ON).
Enables the display.
Disables the display.
Enables the Meter's beeper.
Disables the Meter's beeper.
Remote Operation
Supported SCPI Commands
59