Command
INITiate
DATA
:FEED RDG_STORE, {"CALCulate"|" "}
:FEED?
READ?
SAMPle
:COUNt {<value>|MINimum|MAXimum}
:COUNt? [MINimum|MAXimum]
TRIGger
:COUNt {<value>|MINimum|MAXimum|INFinite}
:COUNt? [MINimum|MAXimum]
:DELay {<seconds>|MINimum|MAXimum}
:AUTO {OFF|ON}
:AUTO?
:DELay? [MINimum|MAXimum]
:SOURce {Bus|IMMediate|EXTernal}
:SOURce?
[1] Bold type denotes default parameter.
SCPI Command Details
All SCPI commands recognized by the Meter are listed in the following sections along
with description and syntax rules.
Table 17. Triggering Command Summary
[1]
Set trigger system to wait-for-trigger
Path to storing readings taken with INITiate
command
Enables/disables storing of INITiate readings
Retrieves state of storing INITiate readings
Set trigger system to wait-for-trigger
Path to sample count
Set number of samples per trigger
Retrieve sample count
Path to trigger control
Set the number of triggers Meter will accept
Retrieve the number of tiggers Meter will accept
Set the delay between trigger and measurement
Retreive present trigger delay value
[1]
Set meter's trigger source
Retrieve the meter's present trigger source
Description
Disable/Enable automatic trigger delay
Retrieve automatic trigger delay status
Remote Operation
Supported SCPI Commands
37