Table 33. Execution Error Codes and Descriptions (cont.)
Error Code
Error Text
-230
Data stale
-242
Date timeout
-243
Second function invalid
-318
Can't save MAC
-319
Can't save LAN
configuration
-320
Can't save non-volatile
configuration
-330
Self-test failed
-350
Too many errors
-410
Query interrupted
-420
Query
UNTERMINATED
-430
Query DEADLOCKED
-440
Query
UNTERMINATED after
indefinite response
-501
RTC Time
-502
RTC Data
31
mA AC/DC selected
while REAR inputs
selected.
Description
A FETCh? command was received but internal reading
memory was empty. The reading retrieved may be invalid.
Timed out while taking data.
Secondary function is not enabled.
Could not save MAC address.
Could not save network configuration.
Could not save non-volatile configuration.
The Meter's complete self-test failed from the remote interface
(*TST? command). In addition to this error, more specific self-
test errors are also reported.
The error queue is full because more than 16 errors have
occurred. No additional errors are stored until you remove
errors from the queue. The error queue is cleared when power
has been off, or after a *CLS (clear status) command has been
executed.
A command was received which sends data to the output
buffer, but the output buffer contained data from a previous
command (the previous data is not overwritten). The output
buffer is cleared when power has been off, or after a *RST
(reset) command has been executed.
The Meter was addressed to talk (i.e., to send data over the
interface) but a command has not been received which sends
data to the output buffer. For example, a CONFigure command
(which does not generate a response) may have been
executed, followed by an attempt to read data from the remote
interface.
A command was received that generated too much data to fit
in the output buffer, and the input buffer is also full. Command
execution continues but all data is lost.
The *IDN? command must be the last query command within a
command string. Example: *IDN?;:SYST:VERS?
Error setting the RTC/System time.
Error setting the RTC/System date.
8842A emulation error. The Front/Rear input switch was set to
REAR while ac/dc current was selected via a remote command
or ac/dc current was selected and the Front/Rear switch was
switched to REAR.