Detailed Specifications - Fluke TiS10 User Manual

Performance series thermal imagers
Hide thumbs Also See for TiS10:
Table of Contents

Advertisement

Korea (KCC) ....................................................... Class A Equipment (Industrial Broadcasting & Communication Equipment)
USA (FCC) ......................................................... 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Vibration ................................................................ 2 G, IEC 68-2-6
Shock ..................................................................... 25 G, IEC 68-2-29
Drop ....................................................................... 2 m (6.5 ft) with standard lens
Size (H x W x L) ..................................................... 26.7 cm x 10.1 cm x 14.5 cm (10.5 in x 4.0 in x 5.7 in)
Weight (includes battery)
TiS10, TiS20, TiS40, TiS50, TiS60 ..................... 0.72 kg (1.6 lb)
TiS45, TiS55, TiS65 ........................................... 0.77 kg (1.7 lb)
Enclosure Rating .................................................. IP54
Warranty ................................................................ 2 years
Calibration Cycle .................................................. 2 years (assumes normal operation and normal aging)
Supported Languages .......................................... Czech, Dutch, English, Finnish, French, German, Hungarian, Italian, Japanese, Korean, Polish, Portuguese,

Detailed Specifications

Temperature Measurements
Temperature Range (not calibrated below -10 °C)
TiS10 .............................................................. -20 °C to +250 °C (-4 °F to +482 °F)
TiS20, TiS40, TiS45 ....................................... -20 °C to +350 °C (-4 °F to +662 °F)
TiS50, TiS55 ................................................... -20 °C to +450 °C (-4 °F to +842 °F)
TiS60, TiS65 ................................................... -20 °C to +550 °C (-4 °F to +1022 °F)
Accuracy ............................................................. ±2 °C or 2 % (whichever is greater) at 25 °C ambient
On-screen emissivity correction ......................... 1 % to 100 %
On-screen reflected background
temperature compensation ................................. yes, default is 22 °C
On-screen transmission correction
TiS45, TiS55, TiS65 (manual focus) ............... 1 % to 100 %
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller
or user should take notice of it. This equipment is intended for use in business environments and not to
be used in homes.
Russian, Simplified Chinese, Spanish, Swedish, Traditional Chinese, and Turkish
Performance Series Thermal Imagers
Detailed Specifications
39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tis20Tis45Tis40Tis50Tis55Tis60 ... Show all

Table of Contents