Power On And Off; Controls For Image Capture; Laser Pointer - Fluke TiS10 User Manual

Performance series thermal imagers
Hide thumbs Also See for TiS10:
Table of Contents

Advertisement

TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65
Users Manual

Power On and Off

To turn on or turn off the Imager, push and hold
>3 seconds. The Imager has Power Save and Auto Off
features. For more information about how to set these
features, see Settings Menu.
All thermal imagers need sufficient warm-up time
for the most accurate temperature measurements
and best image quality. This time can often vary
by model and by environmental conditions.
Although most imagers are fully warmed up in 3-5
minutes, it is always best to wait a minimum of
10 minutes if the most accurate temperature
measurement is important to your application.
When you move the Imager between
environments with large differences in ambient
temperature, more adjustment time can be
required.
The Imager includes a calibration on-demand feature that
causes a calibration event to occur when you press 
once briefly during operation. This feature provides the
best accuracy and avoids disruption of a timing-sensitive
image capture by the next automatic calibration.
10
Note

Controls for Image Capture

The two-part trigger is located in the standard trigger
for
position for a pistol-grip device. The larger, green trigger is
the primary trigger. The smaller, black trigger is the
secondary trigger.
In normal operation (video is off), the function of the
primary trigger is to capture a thermal image for possible
storage to memory by the user. When video is on, the
primary trigger is the start/stop for video recording.
The secondary trigger operates the laser on supported
models.

Laser Pointer

The TiS45, TiS50, TiS55, TiS60, and TiS65 models
include a laser pointer. The laser pointer is a sighting aid
and is offset from the infrared camera. As a result, it may
not always represent the exact center of the infrared or
visible image.
The laser dot does not appear on an infrared-only image,
but does on visible-only or AutoBlend images. The laser
dot cannot be seen in the visible channel of the IR-Fusion
image if obscured by the center point marker graphic.
Pull the secondary trigger to turn on laser pointer, release
the secondary trigger to turn off the laser pointer.