Troubleshooting - Abbott FreeStyle InsuLinx Owner's Booklet

Hide thumbs Also See for FreeStyle InsuLinx:
Table of Contents

Advertisement

Troubleshooting

1. Test strip
is inserted
in the port
and nothing
happens.
33
What It Means
Test strip is inserted
incorrectly or not fully
inserted into the meter
No batteries are installed
Batteries are installed
incorrectly
Dead batteries
Meter may be plugged
into a computer
Problem with test strip
Problem with meter
Blood or foreign objects
put in the test strip port
What To Do
Insert the test strip as shown .
Install batteries with (+) facing up .
1 . Change batteries . Reset date and time,
if needed .
2 . If problem continues, contact Customer
Service .
Unplug the meter from the computer .
Try a new test strip .
Contact Customer Service .