6 Adjustment

The balance must be adjusted at its place of installation before initial use and at regular
intervals.
A warm-up time of about 4 hours is recommended for stabilisation.
Please use the recommended adjusting weight according to the technical data in the
operating instruction.

6.1 Adjustment of non-verification balances

6.2 Adjustment of verification balances (KERN 573-NM; KB-NM; FKB-M)

Remark: The adjusting is only possible when not being blocked by the adjusting switch.
The switch for this step is located at the bottom of the housing between the two
turning feet.
Switch to the right
Switch to the left
572-SH-e-0951
Adjusting function is released.
This setting has to be selected before starting the adjusting.
Executing the adjusting according to below procedure.
After the adjusting has been completed successfully it is
necessary to turn the adjusting switch to the left for blocking.
Adjusting function is blocked.
After the adjusting has been completed successfully the
adjusting has to be blocked by switching to the left.
The balance is now prepared for the procedure of verification.
After the verification the adjusting switch has to be secured by
a verification mark against access.
7