Error Messages - KERN HFB-BA-e-1212 Operating Instructions Manual

Kern hfb-ba-e-1212 electronic crane scales
Table of Contents

Advertisement

8. Error messages

Error
Description
message
Maximum load exceeded
--ol-
Zeroing range exceeded
Err 4
(normally 4% max)
Err 5
Keyboard error
Value outside the A/D changer
Err 6
range
At an instable weighing value
Err 9
ZERO or TARE button
actuated
Err 17
Tare value out of range
Capacity of rechargeable
Ba lo
battery exhausted
Should other error messages occur, switch balance off and then on again. If the error
message remains, inform manufacturer.
26
Possible causes
 Reduce load
 Check whether the balance has
been damaged
 Overload when zeroing
 Improper operation of the balance
 Weighing plate not installed
 Damaged weighing cell
 Damaged electronics
 Wait for a stable weighing value
 Reduce tare load
 Recharge battery
HFB-BA-e-1212

Advertisement

Table of Contents
loading

Table of Contents