8. Error messages
Error
Description
message
Maximum load exceeded
--ol-
Zeroing range exceeded
Err 4
(normally 4% max)
Err 5
Keyboard error
Value outside the A/D changer
Err 6
range
At an instable weighing value
Err 9
ZERO or TARE button
actuated
Err 17
Tare value out of range
Capacity of rechargeable
Ba lo
battery exhausted
Should other error messages occur, switch balance off and then on again. If the error
message remains, inform manufacturer.
26
Possible causes
Reduce load
Check whether the balance has
been damaged
Overload when zeroing
Improper operation of the balance
Weighing plate not installed
Damaged weighing cell
Damaged electronics
Wait for a stable weighing value
Reduce tare load
Recharge battery