Setting Damping - Seca 644 Instruction Manual

Hide thumbs Also See for 644:
Table of Contents

Advertisement

– You can alter the weight of the patient
using the arrow keys in display steps.
– Once the correct value is set, press the
FUNC key again.
– The pre-Tare function is now activated.
– Mount the scale together with the pa-
tient. Read the weighing result off the
digital display.
– Press an arrow key if you would like to
see the current weight value briefly.
– To return to normal weighing mode,
press the FUNC key.
– To switch off the scale, press the Start
key again.

Setting damping

Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display re-
acts. You can set damping to suit requirements.
– Press the FUNC key and keep it de-
pressed for 2 seconds.
FIL and the current setting appear in
the display:
0 = slight damping
fast determination of weight
1 = average damping
normal determination of weight
2 = strong damping
slow determination of weight
– Press an arrow key if you want to
change the set value.
– Press the FUNC key again to store the
displayed value. The scale is then back
in normal weighing mode.
The settings are retained even once
the scale is switched off.
32

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents