Sony UP-D711MD Instructions For Use Manual page 4

Digital graphic printer
Hide thumbs Also See for UP-D711MD:
Table of Contents

Advertisement

The UP-D711MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
UP-D711MD should assure that it is used in such as environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
4
WARNING
Guidance and manufacturer's declaration-electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
< 5% U
T
(> 95% dip in U
) for 0.5
T
cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(> 95% dip in U
) for 5 sec
T
3 A/m
Compliance level
±6 kV contact
±8 kV air
±1 kV for input/output lines
Not applicable.
Not applicable.
3 A/m
Electromagnetic
environment-guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the UP-
D711MD requires continued
operation during power mains
interruptions, it is recommended
that the UP-D711MD be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at least characteristic of
a typical location in a typical
commercial or hospital
environment.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents