Frequency Counter Sensitivity - Fluke CNX 3000 User Manual

Wireless multimeter
Hide thumbs Also See for CNX 3000:
Table of Contents

Advertisement

Frequency Counter Sensitivity

[1] [2]
Input Range
6 V
60 V
600 V
1000 V
6 V
60 V
600 V
1000 V
60.00 mA
400.0 mA
[1]
Maximum input for specified accuracy = 10X Range or 1000 V.
[2]
Noise at low frequency and amplitude may exceed the frequency accuracy specification.
[3]
Unspecified but usable depending on quality and amplitude of signal.
[4]
In mA and A ranges, frequency measurement is specified to 10 kHz.
Quality Fluke Products Online at:
2 Hz to 45 Hz
45 Hz to 10 kHz
0.5 V
0.6 V
5 V
3.8 V
50 V
36 V
500 V
300 V
0.5 V
0.75 V
4 V
3.8 V
40 V
36 V
500 V
300 V
5 mA
4 mA
5 mA
4 mA
GlobalTestSupply
www.
Typical Sensitivity (RMS Sine Wave)
10 kHz to 20 kHz
1.0 V
4.1 V
39 V
320 V
1.4 V
4.3 V
39 V
320 V
NA
NA
.com
sales@GlobalTestSupply.com
Wireless Multimeter
Specifications
20 kHz to 50 kHz
50 kHz to 100 kHz
2.8 V
Unspecified
5.6 V
9.6 V
50 V
58 V
380 V
NA
4.0 V
Unspecified
6.6 V
13 V
45 V
58 V
380 V
NA
NA
NA
NA
NA
[3]
[3]
41