Comparing Glucose Meter Test Results With Laboratory Results - iHealth BG1 Owner's Manual

Align gluco-monitoring system
Hide thumbs Also See for BG1:
Table of Contents

Advertisement

• Expired or contaminated test strip.
• Meter malfunction.
If you continue to get control solution test results that fall outside of the range printed on the vial,
the meter may not be working properly. Discontinue use and call iHealth Labs Customer Service
at 1-855-816-7705 for assistance.
To purchase additional control solution, call iHealth Labs Customer Service.
NOTE:
• Do not use expired control solution.
• The control solution range printed on the test strip vial is for control solution use only. It is not a
recommended range for your blood glucose level.
COMPARING GLUCOSE METER TEST RESULTS WITH
LABORATORY RESULTS
The iHealth system provides you with whole blood equivalent results. The result you obtain from
your glucose meter may differ somewhat from your laboratory results due to normal variation.
The iHealth system results can be affected by factors and conditions that do not affect laboratory
results in the same way.
To make an accurate comparison between the iHealth system and laboratory results, follow the
guidelines below.
Before the Lab Test
• Perform a control solution test to make sure that the meter is working properly.
• If possible, fast at least eight hours before conducting a comparison test.
• Take the iHealth system with you to the lab.
While at the Lab
Make sure that samples for both tests are taken and tested within 15 minutes of each other.
• Wash your hands before obtaining a blood sample.
• Never use your glucose meter with blood samples collected in a test tube.
• Use fresh capillary blood only.