Operation
Three-point calibration
22
11
Press the <RUN/ENTER> key. Ct3 appears on the display.
ISE
U
pH
S
8
C t
1
Time
Baud
8
2
4
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
12
Thoroughly rinse the electrode with distilled water.
13
Immerse the pH electrode in the third buffer solution.
14
Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
ISE
U
pH
S
1
1
7 1
7
Time
Baud
8
2
4
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
15
When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The probe symbol
shows the evaluation of the current calibration (three-point cal-
ibration).
mol/L
3
mV/pH
% ppm
mg/L
°F
8
%
TP
°C
CalError
TempError
ARng AR
RCL
mol/L
S
1
0
mV/pH
mV
% ppm
mg/L
°F
Time
8
%
Day.Month
TP
°C
Year
CalError
TempError
AutoCal DIN
LoBat
ARng
AR
RCL
AutoCal TEC
ISE
U
pH
mol/L
0
1
1
mV
mV/pH
% ppm
°F
Baud
8
2
4
8
No.
°C
Ident
CalError
Auto Store
ARng
ConCal
pH 315i
mg/L
%
TP
TempError
AR
RCL