When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The probe symbol
shows the evaluation of the current calibration (two-point cali-
bration).
ISE
U
pH
S
-
5 9
1
Time
Baud
s
8
L
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
9
Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
ISE
U
pH
S
6
5 9
1
Time
Baud
8
a
s
Day.Month No.
Year
Ident
AutoCal DIN
Auto Store
LoBat
AutoCal TEC
ConCal
10
Change to the measuring mode with <M>, or proceed to the
three-point calibration with <RUN/ENTER>.
mol/L
4
mV/pH
% ppm
mg/L
°F
0
%
TP
°C
TempError
CalError
ARng
AR
RCL
mol/L
4
mV/pH
mV
% ppm
mg/L
°F
y
%
TP
°C
TempError
CalError
ARng
AR
RCL
Operation
21