5.2

Troubleshooting

Problem
The reading is
extremely low (or
high).
Arm cuff pressure
does not rise.
Arm cuff deflates too
soon.
Cannot measure or
readings are too low
or too high.
The unit loses power
during measurement.
Cause
Arm cuff not applied
correctly.
Movement or talking
during measurement.
Clothing is interfering
with the arm cuff.
The air plug is not
securely inserted into
the main unit.
Air is leaking from the
arm cuff.
The arm cuff is loose.
The arm cuff has not
been inflated
sufficiently.
The batteries are
empty.
5.Handling Errors and Problems
Remedy
Apply the arm cuff
correctly.
Refer to section 3.2.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air
plug is connected
securely.
Refer to section 3.2.
Replace the arm cuff
with a new one.
Refer to Chapter 7.
Apply the cuff correctly
so that it is firmly
wrapped around the
arm.
Refer to section 3.2.
Inflate the cuff so that it
is 30 to 40 mmHg
above your previous
measurement result.
Refer to section 3.4.
Replace the batteries
with new ones.
EN
25

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M2 Basic and is the answer not in the manual?

Questions and answers

Helen Coward
March 10, 2025

can you get more than the last saved reading

2 comments:
Mr. Anderson
March 10, 2025

Yes, the Omron M2 Basic can store and display up to 21 measurements.

This answer is automatically generated

Helen Coward
March 10, 2025

@Mr. Anderson How do you get the other readings. I press on the start button for more than 5 secs and it gives me the last reading - how do I get previous ones?

Table of Contents

Save PDF