Care and maintenance

5. Handling Errors and Problems

5.1

Error Messages

Error Display
Cause
Cuff is under inflated.
Movement during
measurement.
Air plug disconnected.
Arm cuff not applied
correctly.
Clothing is interfering
with the arm cuff.
Air is leaking from the
arm cuff.
The arm cuff was
inflated above 299
mmHg when inflating
the cuff manually.
5.Handling Errors and Problems
Remedy
Carefully read and repeat the
steps listed under section 3.4.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Insert the air plug securely.
Refer to section 3.2.
Apply the arm cuff correctly.
Refer to section 3.2.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.2.
Replace cuff with new one.
Refer to Chapter 7.
Do not inflate the arm cuff
above 299 mmHg.
Refer to section 3.4.
EN
23

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M2 Basic and is the answer not in the manual?

Questions and answers

Helen Coward
March 10, 2025

can you get more than the last saved reading

2 comments:
Mr. Anderson
March 10, 2025

Yes, the Omron M2 Basic can store and display up to 21 measurements.

This answer is automatically generated

Helen Coward
March 10, 2025

@Mr. Anderson How do you get the other readings. I press on the start button for more than 5 secs and it gives me the last reading - how do I get previous ones?

Table of Contents

Save PDF