Operation

Electric shock hazard:
Contact with live circuits could result in severe injury or death.
1. Refer to the Settings Table. Set the selector to the proper setting, press SELECT (when instructed to
do so), and connect the test leads to the meter.
2. Refer to "Typical Measurements" for specific measurement instructions.
3. Test the unit on a known functioning circuit or component.
• If the unit does not function as expected on a known functioning circuit, replace the battery and/
or fuses.
• If the unit still does not function as expected, call Greenlee for technical assistance
at 800-435-0786.
4. Take the reading from the circuit or component to be tested.

Settings Table

To measure this
characteristic ...
All Models
Voltage
(1000 V max)
Resistance
Continuity*
Diode
Voltage (600 mV max)**
Current (8 A max)**
Current (600 mA max)**
Current (6000 µA max)**
This table continues on the next page.
12
Set the
These icons will
selector to this
appear on the
symbol ...
Ω
and press
diod (momentarily)
SELECT
mV
A mA
A mA
mA,
µA
µA,
Connect
the red
display ...
lead to ...
and V
and V
and Ω
and V
mV
A,
or
mA µA
or
mA µA
or
Connect
the black
lead to ...
ΩV
COM
ΩV
COM
ΩV
COM
A
COM
COM
COM