SD CodeFree™ BLOOD GLUCOSE MONITORING SYSTEM
accordance with local guidelines.
4) Control solution
If you are a medical professional or have the
control solution in your possession please note
the following.
Keep the SD Control Solution in a cool dry
environment 8-30°C (46-86°F) .
Do not refrigerate or freeze.
Do not use SD Control Solution that has
passed its expiration date.
The SD Control Solution can be used for 3
months after opening the container. Do
not use after this time. Write the opened
date on the SD Control Solution container
when first opened.
No reconstitution or dilution is necessary.
Wipe the container tip clean and reseal the
container tightly after each use.

4. Cleaning the Lancing Device

Clean the outside of the SD lancing device
regularly with 70% isopropyl (rubbing) alcohol.
Do not place the entire device under water. Do
not use bleach. At least once a week, disinfect
the removable comfort cap after cleaning by
placing it in 70% rubbing alcohol for 10 minutes.
Allow the cap to air-dry after disinfecting.
64