SD CodeFree™ BLOOD GLUCOSE MONITORING SYSTEM
strips have been damaged.
The meter has been dropped or damaged.
Before you begin ;
Use only SD Control Solution.
Check the expiration date on the control
solution container. Record the opening
date on the container label. Do not use after
expiration or discard date (date opened plus
three months), whichever comes first.
Control solution, meter, and test strips should
be at room temperature 18-30°C (64-86°F)
before testing with control solution.
Shake the container, discard the first drop of
control solution, and wipe off the tip to ensure
a proper sample and an accurate result.
Store control solution tightly closed at
temperatures between 8-30°C (46-86°F). Do
not refrigerate.
CAUTION
32
Do not swallow control solution - it
is not for human consumption.
Do not apply control solution to
the skin or eyes as it may cause
irritation.