Invacare Kite Service Manual page 73

Hide thumbs Also See for Kite:
Table of Contents

Advertisement

®
Invacare
- Kite
·
Use a 6 mm Allen wrench and a 13 mm socket wrench to loosen and remove the four
hexagon socket head screws (1).
·
Remove the screws together with the washers.
·
Pull the motor/gear box unit (3) forwards out of its holder (2). Take into account the heavy
weight of the unit.
Refitting the motor/transmission unit:
·
Replace the defective components.
·
Before installing a new motor / gearbox unit, make sure the sealing ring is correctly mounted.
The sealing ring securely attaches the motor and the gearbox. The ring must be installed such
that the self-locking nut is located on the outside.
·
If the sealing ring is incorrectly installed, rotate it around as described in chapter 8.10.2.
· Check the serial number of the gearbox housing.
Up to the serial number 20110400xx, the gearbox housing is
44.5 mm wide: '
The compensation sheet (see right) must be installed.
Up to the serial number 20110400xx, the gearbox housing is
49.5 mm wide:
The compensation sheet is not necessary.
·
Install the motor / gearbox unit in reverse order.
·
Check the orientation of the motor and adjust it.
Make sure that the motor cable cannot be pinched or bent, and that it is not exposed to
chafing. The coupling lever must be able to move to the positions "Driving" as well as
"Pushing".
·
Tighten the self-locking nut of the sealing ring to a torque of 4 Nm.
·
Connect the motor cable plug.
·
Reinstall all components that were removed beforehand in reverse order.