Compatibility
Command
OVSET <n>
This command programs the overvoltage protection. See Table
8-3 for the programming range of this command.
Initial condition: MAX
POL 0 | 1
Only applies to units with Option 760. This command sets the
polarity of the output relays to either normal (1) or inverted (0).
Initial condition: POL 1
PON 0 | 1
This command enables (1) or disables (0) SRQ at power-on.
Initial condition: last programmed value
RELAY 0 | 1
Only applies to units with Relay Option 760. This command
opens (0) or closes (1) the output relays without affecting the
programmed output state of the unit.
Initial condition: RELAY 1
RLYPON 0 | 1
Only applies to units with Relay Option 760. This command
opens (0) or closes (1) the output relays at power-on without
affecting the programmed output state of the unit.
Initial condition: RLYPON 1
ROM?
This command returns the ROM version of the dc source.
Data Representation: AAA AAA
RST
This command resets the dc source if the output is disabled by
the output protection circuits.
SENS:CURR
This command sets the current measurement range of the dc
:RANG <n>
source. See Table 8-3 for the programming range.
Initial condition: MAX
SENS:PROT
This command enables/disables the open sense detect feature.
:STAT 0 | 1
Initial condition: 1
SENS:SWE
This command defines the number of data points in a
:POIN <n>
measurement sweep.
Initial condition: 32
SENS:SWE
This command defines the time period between measurement
:TINT <n>
samples.
Initial condition: 15.6 s.
SRQ 0 | 1
These commands enable or disable the dc source's ability to
request service from the controller for fault conditions.
UNMASK defines which conditions are defined as faults.
Initial condition: SRQ 0
STS?
This command reads the contents of the status register, which
maintains the present status of the dc source.
Data Representation: ZZZZD
SYST:LANG
This command causes the alternate language to become active
and to be stored in nonvolatile memory. In this case, the
commands are equivalent. After being shut off, the dc source
will resume in the last-selected language when power is
restored. The parameter must be either COMP or SCPI.
Table F-2. COMPatibility Commands (continued)
Description
Compatibility Language - F
Similar SCPI
Command
VOLT:PROT
OUTP:REL:POL 0|1
PSC 0 | 1
OUTP:REL 0 | 1
RCL 0
*IDN?
OUTP:PROT:CLE
SENS:CURR:RANG
SENS:PROT:STAT
SENS:SWE:POIN
SENS:SWE:TINT
*SRE
STAT:OPER:COND?
STAT:QUES:COND?
*ESE?
SYST:LANG