Additional Commands

The following commands are added to the language dictionary to provide limited Fluke PM2811 Power
Supply compatibility. These comands will only work when the SYSTem LANGuage is set to SCPI.

INSTrument:STATe

This command enables or disables the dc source output. The state of a disabled output is a condition of
zero output voltage and a model-dependent minimum source current (see *RST).
Command Syntax INSTrument:STATe <bool>
Parameters 0 | OFF | 1 | ON
*RST Value 0
Examples INST:STAT 1
Query Syntax INSTrument[:STATe]?
Returned Parameters <NR1> 0 or 1
Equivalent SCPI Command OUTPut[:STATe]

OUTPut:PROTection:TRIPped?

This query reports if any OV (overvoltage), OCP (overcurrent protection), OT (over temperature), or RI
(remote inhibit) status bit is set. 0 = bit is not set; 1 = bit is set.
Query Syntax OUTPut:PROTection:TRIPped?
Returned Parameters <NR1> 0 or 1
Examples OUTP:PROT:TRIP?
Equivalent SCPI Command STATus:QUEStionable:CONDition?

CURRent:LIMit:HIGH?

This query returns the maximum rated current of the instrument.
Query Syntax [SOURce:]CURRent:LIMit:HIGH?
Returned Parameters <NRf>
Examples CURR:LIM:HIGH?
Equivalent SCPI Command [SOURce:]CURRent? MAX

CURRent:LIMit:LOW?

This query returns the minimum rated current of the instrument.
Query Syntax [SOURce:]CURRent:LIMit:LOW?
Returned Parameters <NRf>
Examples CURR:LIM:LOW?
Equivalent SCPI Command [SOURce:]CURRent? MIN