Parameter
Measurement Time
1
(includes 30 ms
data
acquisition time and 20 ms
data processing overhead)
Command Processing Time
(time for output to begin to
change following receipt of
digital data)
Savable Instrument States
(applies only in SCPI mode)
GPIB Interface Capabilities
INH/FLT Characteristics
Digital I/O Characteristics
Isolation to Ground
(Maximum from either output
terminal to chassis)
Mains Input Ratings:
(at full load from 47–63 Hz)
1
May be reduced by changing the default conditions of 2048 data points but measurement uncertainty due to noise will increase.
2
Compatibility language is used to program the Agilent 663xA Series power supplies. Not all Agilent 66111A/66311B
capabilities are available when using the Compatibility language. Also, Compatibility language is not available for Agilent
66309B/D units.
Table A-2. Supplemental Characteristics (continued)
voltage or current
Language:
Interface:
Maximum ratings:
FLT Terminals:
INH Terminals:
Maximum ratings:
Digital OUT Port 0,1,2
(open collector)
Digital IN Port 2:
(internal pull-up)
100 Vac (87-106 Vac):
115 Vac (104-127 Vac):
220 Vac (191-233 Vac):
230 Vac (207-253 Vac):
Agilent 66111A
Agilent 66311B/D
50 ms average
4 ms average
4 (in locations 0 to 3)
SCPI or COMPatibility
AH1, C0, DC1, DT1, E1,
L4, PP0, RL1, SH1, SR1, T6
16.5 Vdc between terminals 1 and 2; 3 and 4; and
from terminals 1 or 2 to chassis ground
Low-level output current = 1.25 mA max.
Low-level output voltage = 0.5 V max.
Low-level input voltage = 0.8 V max.
High-level input voltage = 2 V min.
Low-level input current = 1 mA
Pulse width = 100 µs min.
Time delay = 4 ms typical
same as INH/FLT Characteristics
Output leakage @ 16V = 0.1 mA (ports 0,1)
Output leakage @ 5V = 0.1 mA (ports 0,1)
= 0.25 mA (port 2)
Low-level output sink current @ 0.5 V = 4 mA
Low-level output sink current @ 1 V = 50 mA
Low-level input current @ 0.4 V = 1.25 mA
High-level input current @ 5 V = 0.25 mA
Low-level input voltage = 0.8 V max.
High level input voltage = 2.0 V min.
50 Vdc
1.7 A, 125 W
1.5 A, 125 W
0.8 A, 125 W
0.75A, 125 W
Specifications - A
Agilent 66309B/D
2
= 12.5 mA (port 2)
2 A, 170 W
1.7 A, 170 W
0.92 A, 170 W
0.85A, 170 W