Call bra tlon
3457A
5-9. In the event that the security code is unknown, the security feature can
be
disabled t o permit a new
code t o be entered. To disable the security feature, complete the following steps.
a. Remove all power from the 3457A (line cord and external inputs) and turn the unit upside-down.
b. Loosen the six captive screws in the bottom cover.
( I t is not necessary to remove these screws from the bot-
tom cover).
c. Return the unit to an upright position and remove the top cover
d. Remove the
NORM
/
DSA
/
KEY
jumper (JM532) from the
NORM
position and place it in the
K E Y
position.
( J M J 3 2 is located on the A l logic board directly behind the front panel display).
e. Reconnect the power and turn the instrument
ON.
f. Access the
SECURE
command.
(Press the
Blue
SHIP?' key then the HECALI,
/ S
configuration
key. Use the
5.
scroll key t o display the S E C U R E command).
g. Enter the number 0, the delimiter
(
,
)
and the security number you wish t o use.
h. Press the
ENT
key.
i. Disconnect power and return jumper JM532 to the
NORM
position.
j, Replace the top cover, tighten the retaining screws and reconnect the power. The instrument will
now respond to the security code just entered.
When jtirnper JM532 is in the KEY position, fhc sccririty feafrrre is disabled. I f is
possible to calibrate /he instrrrrnenl wilhorrt enlering a ~ecrrrity number under f h e ~ e
conditions. I f a new stwrrity nrrrnber is nof twtertld whilc fhtl jrrrnper is in the KEY
position, the original nrrrnber will again he in e\jec/ whcn jrrrnpcr JM532 is retrrrn-
ed to the NORM position.
5-10. Calibration Number (CALNUM?)
5-1 1. The
CALNUM?
command provides you with a method of monitoring the number of calibrations per-
formed. By using this command, you can determine whether un-authorized calibrations have occurred. The
instrument will always respond the
CALNUM
query command by displaying the number of calibration
RAM entries completed since factory initialization. factory initialization. The maximum calibration num-
ber which can be stored is 32767. After reaching the maximum number, the counter resets to 0 and begins
again. Please note that the calibration number is increased by one for each calibration point. A complete
calibration increases the calibration number by several digits.
5-12. The procedure for reading the Calibration Number is as follows:
a. Access the
CALNUM?
command.
(Press the Blue S H I F T key then the OFFSET C O M P
/ C
configuration key.
Use the
5.
scroll key t o display the CALNUM? command).
b. Press the
ENT
key.