H P 44492A Reed Relay - HP 3457A Service Manual

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

Operation
3457A
Make the measuren~ent
-
Press the
SINGLE
/
F
configuration key to start the ~ncasurenient sequence.
The instrument will read each channel listed and store the readings i n memory.
Recall the readings
-
Press the Reading Memory
RECALL
/
S
hey then the
ENT
/
LAST ENTRY
key.
The reading displayed is the last reading taken. The readings are shifted into mcmory so that the first
reading taken is in the highest number memory location. The last reading taken is in memory location 1.
Use the ?/AUTO and
HOLD
scroll keys t o view the measurement readings. Use the
+/TEST
and
+/RESET
display keys to read the total display contents.
T o repeat the same set of measurements
-
Press the
CONFIG
/
R
configuration key then the
ENT
/
LAST ENTRY
key to clear the reading memory. Press the
SINGLE
/
F
conl'iguration key t o start
the new reading sequence.
3-45. HP 44492A Reed Relay Multiplexer (Manual Operation)
3-46. The HP 44492A Reed Relay Multiplexer has len 2-wire input channels nurnbered 0 through 9. This
multiplexer is wed t o make voltage, resistance and frecl~~ency or pesiod measurenlents. It is not designed t o
make 4-wire ohms or current measurements.
3-47. Measurement Selection. Determine the type of measurement you wish to make (Voltage, Resistance
or Frequency/Period) and set the instrument for that measi~remenl as described in lhe previous paragraphs.
3-48. Input Selection. T o use the Reed Relay Multiplexer, the instrument must be set t o the Rear/Scanner
input.
Select the Rear/Scanner input
-
Press the
TERM
/
M
configuration key. Use the
?/AUTO
or
HOLD
scroll key t o display
REAR::!
or
SCANNER::! and
press the
ENT
/
LAST ENTRY
key.
3-49. Channel Selection. Only one input channel can be selected at a time. When an input channel is
closed, the channel previously closed is automatically opened.
T o select an input channel
-
Press the
CHAN
/
N
configuration key, enter the number of the channel t o
be closed and press the
ENT
/
LAST ENTRY
key.
T o open all channels
-
Press the Blue
SHIFT
key then the
OFFSET COMP
/
C
key. Use the
J/HOLD
scroll key t o display
CRESET
and press the
ENT
/
LAST ENTRY
key.
3-50. Scanning. The instr~~nlent can be set t o automatically take a number of readings and store them in
memory. This makes it possible t o measure some or all of the input channels and recall the measurements
taken. The following is a simple procedure to measure inputs connected t o channels 0 through 9 of the
HP
44492A Reed Relay Multiplexer.
Determine the type of measurement you wish to make (Voltage, ?-Wire Resistance. Frequency/Period)
and set the instrument for that measurement as described in the previous paragraphs.
Select the Rear/Scanner input
-
Press the
TERM
/
M configuration key. Use the ?/AUTO or
HOLD
scroll key t o display
REAR::!
or
SCANNER:2
and press the
ENT
/
LAST ENTRY
key.
Set the Trigger to Hold
-
Press the Blue
SHIFT
key then the
STORE
/
T
configuration key. Use the
HOLD
scroll key to display the
TRIG
command. Press the
+/RESET
display hey once t o hold the com-
mand. Use the
?/AUTO
or
HOLD
scroll key to display
HOLD:4
and press the
ENT
/
LAST ENTRY
key.
Set the number of readings to match the number of measurements to be made (in this case 10)
-
Press
the
NDRGS
/ L
config~lration key, enter the n~rmber 10 and press the
ENT
/
LAST ENTRY
key.

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents