Operational Verification Tests
4-Wire Ohms Operational Verification Tests Cont'd
Frequency Counter Operational Verification Tests
Figure 4-1 9. Frequency Counter Operational Verification Test Connections
4-88. Frequency Counter
-
Accuracy Test
4-89. Equipment Required. A
Frequency Source capable of providing a 20 Hz
(+
.01'?!1) and a I M H z
(+
.003'70) sine-wave signal is required f o r this procedure.
Table 4-28. 4-Wire Ohms Gain Test Limits
1 y e a r
l i m i t s
1.
Set the HP 3457A t o the Frequency Function (FREQ).
3457A
S e t
U p
OHMF
OHMF
OHMF
3457A
I
npu
t
3
Kohm
3 0
Kohm
300
Kohm
H i g h
3 . 0 0 0 1 5 7
Kohm
3 0 . 0 0 1 5 7
Kohm
3 0 0 . 0 1 5 8
Kohm
90
d a y l i m i t s
2. Set the Frequency Standard f o r a 1 volt, 20 Hz sine-wave o i ~ t p u t signal and connect it t o the HI
and LO input terminals of the HP 3457A.
3457A
Range
3
Kohm
30
Kohrn
300
Kohm
Low
2 . 9 9 9 8 4 3
Kohm
2 9 . 9 9 8 4 4
Kohm
2 9 9 . 9 8 4 2
Kohm
H i g h
3 . 0 0 0 1 1 2
Kohm
3 0 . 0 0 1 1 2
Kohm
3 0 0 . 0 1 2 8
Kohm
3.
Use the Test Record t o record the frequency readings at 20 Hz and I MHz.
Low
2.999888
Kohm
29.99888
Kohm
299.9872
Kohm
4.
If either of the readings are beyond the limits specified, the instrument should be calibrated.
Refer t o Section V of this manual f o r calibration procedures.