ERROR MESSAGE LIST

Error Message
CARD ERROR
NO CARD
UNFORMATTED CARD
CARD FULL
NO CAMERA HEAD
NO PICTURE
BAD PICTURE
CANNOT OPEN FILE
CAMERA DAMAGED
xxxx
NEED A SCALE ON
TIFF ONLY
SAVE THE PICTURE
RESOLUTION ERROR
SINGLE FRAME ONLY
SET TO 1X AGAIN
44
Mode
Warning Details
REC
PL AY
Cannot record, play or
erase pictures.
CF card is not inserted.
CF card needs
formatting.
CF card is full and does
not have remaining
capacity.
Camera head is not
connected.
No picture is recorded so
playback is impossible.
Playback of selected frame
failed. Other operations are
accepted.
Picture file cannot be
opened with the DP20.
The DP20 may be failed.
Measurement cannot be
started because the
scale display is OFF.
Measurement cannot be
started because the pic-
ture format is not TIFF.
Number of measurements
have reached the maxi-
mum limit so it is neces-
sary to save the picture.
Measurements cannot be
started because the display
resolution is inappropriate.
Measurements cannot be
started on the index display.
During measurement,
zoom shooting is not
X
possible.
Treatment
Insert the CF card properly. If the warning recurs,
format the CF card. Note that formatting results
in erasing all of the picture and document files
in the CF card. If formatting is impossible, the CF
card is unusable.
Insert a CF card.
Insert a CF card.
Format the CF card.
Use another CF card, erase unwanted pictures,
or download existing pictures to the PC then
erase all pictures in the CF card.
Connect the camera head properly as follows:
First set the main switch of the control box to "
(OFF), connect the camera head, and set the main
switch to " " (ON) again.
If the camera head is connected while the control
box is kept ON, the camera head will not be
recognized.
Picture playback in PLAY mode is possible even
when the camera head is not connected.
Insert a CF card containing recorded pictures.
Insert the CF card properly. If the warning recurs,
format the CF card. Note that formatting results
in erasing all of the picture and document files
in the CF card. If formatting is impossible, the CF
card is unusable.
A picture recorded using another digital camera
should be played on the digital camera used for
recording it.
Unplug and plug the AC adapter again. If the
warning recurs, note the error code and contact
Olympus.
The error code refers to the 4-digit number shown
in place of "xxxx".
Set the scale display ON before measurements.
Measurements of played picture reflect the scale
display set in recording, but it is impossible if the
picture has been displayed with the scale display
set to OFF.
Measurement of played picture is possible only
when it is recorded in the TIFF format
The maximum number of measurements per pic-
ture is 10. The picture should be saved after it.
If "Scale display range" is set to LIVE (Display
only on the display screen.), the measurement
results are not saved.
Set to a measurement-compatible display resolu-
tion (UXGA 1600 x 1200 or SVGA 800 x 600).
Display only the measurement-target picture frame.
Shoot with the 1X display.
: May be displayed.
: Depends on picture.
Ref.
Page
11,24
11
24
30
"
8
29
11,24
9
21
34
27
26
27
27
: Not displayed.