Specifications - Nikon eclipse e100 Instructions Manual

Table of Contents

Advertisement

Specifications

(1) Model Name:
(2) Optical System:
(3) Focusing Mechanism:
(4) Stage:
(5) Revolving Nosepiece:
(6) Electrical Specifications
Lamp Rating:
Output Rating:
Input Rating:
ECLIPSE E100 (Microscope basic unit)
CFI optical system (infinity - corrected CF optical system)
2nd objective lens focal length : f = 200 mm
Built-in diascopic illumination system (Simplified Kohler's illumination system)
Fine focus knob graduation:
Fine focus knob travel:
Coarse focus knob travel:
Stage vertical movable range: 2 mm upward and 20 mm downward from the
Stroke:
X axis: 78 mm
Y axis: 40 mm
4-hole fixed type
Halogen lamp 6V-20W (PHILIPS 7388 or OSRAM HLX 64250)
6V 3.3A max.
Model for 100, 110, and 120 V AC Areas
Input voltage:
Frequency:
Voltage fluctuation:
Rated current:
Fuse rating:
Power Cord:
2 µm/graduation
0.2 mm up or down / revolution
about 37.7 mm up or down / revolution
focal plane.
Select from 100 V, 110 V or 120 V AC by relocating the
fuse holder in the AC inlet.
50 - 60 Hz
±10%
0.5 A
250 V, 1A, time-lag low-breaking type, 5 x 20 miniature
fuse x 2
Use only the following power supply cord. Using the
wrong power cord could result in danger or fire. The
protection Class I equipment should be connected to
PE (protective earth) terminal.
UL listed detachable power cord set. 3-conductor
grounding type SVT, No.18 AWG, 3 m long
maximum, rated at 125 V AC minimum.
41