2 Inspection and assembly ■ Inspection and assembly Inspect delivery for complete supply immediately on unpacking all components. Components supplied, depending on balance type • Weighing pan M, A • Pan holder • Protective ring for SM and A balances •...
■ 3 Firmware and serial number Firmware and serial number After a reconnection of the balance to the mains and switching on for the first time the serial number as well as the firmware will be showed in the upper display. Display Remark 4700051...
5 Below-balance weighing ■ Below-balance weighing Objects which, because of their size or shape, cannot be put on the scale, can be weighed by means of be- low-balance weighing. Proceed as follows: • Turn off the balance. • Remove the weighing pan and the pan holder than turn the balance up side down. •...
Data cable DB9 Male / DB9 Female (PC), 1.5m 350-8672 Data cable DB9 Male / DB25 Male (Printer), 1.5m 350-8673 Slide-in modules Article number Precisa BUS converter 350-8664 Additional SmartBox® Applications, Precisa BUS accessories, further special accessories and options on de- mand.
Anti-theft encoding ......21 15.3.2 Units..........43 10.3.9 Key tone ........22 15.4 Working with differential weighing..44 10.3.10 Precisa BUS ........22 15.4.1 Mode, measurement recording ..45 10.3.11 Language function ......22 15.4.2 Measurement ........46 10.3.12 Backlight ........22 15.4.3...
Page 11
Contents ■ (MUM, Multiuser Memory) ..64 15.4.6 Report ..........48 15.4.7 Remote control commands ....49 20.1 Activating a user ......64 Minimum original weight 20.2 Creating a new user profile ....64 ("MIN.-WEIGHT") ....50 20.3 Changing the password and password protection ......65 16.1 Explanatory notes on the minimum 20.4...
Virtually all models can be delivered in weights and measures approved. The most important basic production features of the Precisa balances include: • Simple-to-use 10-key multifunction control panel •...
8 Safety ■ Safety Representations and symbols Important safety instructions are highlighted with the appropriate symbol: DANGER Warning of a possible danger which can lead to death or to serious injuries. CAUTION Warning of a possibly dangerous situation which can lead to less severe injuries or damage. NOTE Tips and important rules on the correct operation of the balance.
■ 9 Set up Set up Unpacking the balance The balances are delivered in an environmentally-friendly package, specifically developed for this precision instrument, which provides optimum protection for the balance during transportation. NOTE Retain the original packaging in order to avoid transportation damages when shipping or transporting the balance and to allow the balance to be stored in the best conditions if it is out of operation for an extended period.
9 Set up ■ Choosing a suitable location The balance location must be chosen in such a way as to guarantee perfect operation of your balance, so that the allowable ambient conditions and prerequisites are met and maintained: • Put the balance on a solid, firm and preferably vibration-proof, horizontal base •...
■ 9 Set up NOTE The balance must be calibrated on initial installation and after every relocation. If you work in accordance with “Good Laboratory Practice GLP” observe the prescribed intervals between calibrations (adjustments). Calibration is effected in the configuration menu. Depending on the balance model, this may be done exter- nally, internally or automatically (see chapter 10.3.4 "Calibration functions"...
10 Operation ■ Operation The balance has two main menus available: the configuration menu and the application menu. The basic program of the balance is defined in the configuration menu. With this, you can either work with the basic configuration programmed ex-works, or define and store a user configuration adapted to your spe- cific needs.
■ 10 Operation NOTE For the operation of the «T», « » and «PRINT» see chapter 10.4 "Special operating keys". 10.1.2 Operation in the programming mode Keys Designation Function in programming mode «A», «B» • Changes within the menu level «...
10 Operation ■ Menu Definable functions KEY TONE Activation of the keypad sound BUS OFF Enables disables Precisa BUS accessories LANGUAGE Display language (E, G, F) BACKLIGHT Display backlight DRAFT SHIELD Settings of automatic draft shield, if installed • The settings in the sub-paths pre-programmed at the factory are printed here in bold.
10 Operation ■ DEF. 0.000 g Calibration weight for EXT.-DEF. mode AUTOCAL. TIME/TEMP. Auto-calibration on time and temperature TEMPERATURE Auto-calibration on temperature TIME Auto-calibration on time AUTOCAL.-TIME Time for auto-calibration For calibration of the balance see chapter 9.7 "Calibration of the balance" and see chapter 22.1 "Calibra- tion".
■ 10 Operation „AUTO-STANDBY“ The Auto-Standby mode turns off the balance automatically, if: • the balance is tared and has shown “Zero” for at least 5 minutes • the balance has received no remote control command via the interface for at least 5 minutes, •...
10 Operation ■ 10.3.8 Anti-theft encoding The balance can be protected against theft by using a freely selectable, four-digit numerical code: • If the anti-theft code is deactivated, the instrument can be re-started and operated after a power outage without having to enter a code. •...
• BUS ON/OFF Enables the Precisa BUS functionality, mandatory if Precisa BUS accessories are connected To connect any Precisa BUS accessories switch the BUS to ON, else switch it OFF to have a normal RS232 communication. 10.3.11Language function • LANGUAGE...
10 Operation ■ SENSITIVITY Sensor activation range MEDIUM HIGH SENSOR R. OPEN LEFT Open / close left door OPEN RIGHT Open / close right door TARE Tare PRINT Printout no function SENSOR L. OPEN LEFT Open / close left door OPEN RIGHT Open / close right door TARE...
■ 10 Operation • Print out a balance status – Ensure that the balance is in the Weighing mode – Hold down «PRINT» until „PRINT STATUS” is displayed – Release «PRINT» – The balance status will be printed out. • Print out the application-setup –...
10 Operation ■ 10.5.2 Selecting an application program • SELECT APPLICATION SET APP. normal weighing UNITS different weight units COUNT Piece counting PERCENT Percent weighing CALCULATOR Conversions PAPER Determine paper weight (in g / square cm) NET-TOTAL Add weighing results with intermediate tare Add weighing results without intermediate tare ANIMAL Animal weighing...
■ 11 Application Application 11.1 Units • SETUP APPLICATION UNITS UNIT-2 Milligram ----- ----- not active UNIT-3 Grain ----- ----- not active UNIT-4 Carat ----- ---- not active Assigning the function keys: Function «g» Show measurement in unit 1, e.g. Gram «mg»...
11 Application ■ Working without Tare value: Display Step 0,0000 « » Press until „COUNT“ appears. Count 12,1596 apply (or remove) e.g 25 pieces is re The weight calculated in pieces (PCS) and displayed. « » The reference quantity is set. «PRINT»...
■ 11 Application Display Step 10,0000 «A» „SET“ is flashing. The point of reference is measured 30,0000 apply e.g. 20.0000g 130,000 The weight is recalculated in percent (%) and displayed. «A» The reference quantity is set. All subsequent measurements will now be shown as percentages of the reference weight so defined. «PRINT»...
11 Application ■ Display Step 18,166 «A», «B» The measurement is recalculated accordingly. « », « » name1 NAME2 NAME3 NAME4 «PRINT» The calculated measurement is printed with the set printer text by pressing the key. 11.5 Paper The setting up of the „PAPER” program is similar to that for the calculator (see chapter 11.4 "Calculator"). With the aid of this program you can convert and display the weights of paper samples of standard sizes di- rectly in “gram per square metre”.
■ 11 Application Display Step + 0,0000 « » Display the individual components. 1= 3.4770 9 + 0,0000 «esc» Exit the info display. sto 2 Clear the measurements: + 0,0000 Hold down the key until the acoustic signal sounds and the «...
11 Application ■ Display Step + 8,58962 «esc» Exit the info display. sto 2 Clear the measurements: + 8,58962 Hold down the key until the acoustic signal sounds and the « » component counter is reset. sto 2 + 8,58962 Measurements cleared, balance is ready for a new measurement.
■ 12 Statistics Statistics • SET STATISTICS STATISTICS MODE Statistics program off STATISTICS Statistics only RECORDER Data storage only STAT./RECORDER Statistics and storage COUNT Number of values to be stored automatically (1..500). RECORDING MANUAL With «STO i» function key TIMEBASE on a time basis LOADCHANGE after every weight change...
Page 35
12 Statistics ■ Display Step + 3,4770 «A» Record a stable measurement. sto 0 + 3,4785 «A» Record a second measurement. sto 1 + 3,4775 «Ai» Record a third measurement. sto 2 Retrieve the parameters: + 0,0000 Show info display. «...
■ 13 Check-Weighing Check-Weighing • SET CHECK +/- CHECK +/- MODE ON/OFF Switch application on/off NOM. 100.000 g Enter nominal weight 120.000 g Define over limit 80.000 g Define under limit With the aid of the „CHECK +/-“ program you can check each measurement for its agreement with a defined reference-value plus/minus allowable deviations.
■ 14 Density ("DENSITY") Density ("DENSITY") 14.1 Explanatory notes on density determination 14.1.1 Methods of determination You can use the "DENSITY" workflow to determine the density of solids and liquids. You can choose between different methods of determination: • "MODE LIQUID" : Density of a liquid (only with density determination kit 350-8515) •...
14 Density ("DENSITY") ■ 14.3 Configuring the density determination • SETUP APPLICATION DENSITY MODE SOLID ON BOTTOM Solids with a vessel on the weighing pan MODE SOLID IN AIR Solids MODE LIQUID Liquids MODE SOLID POROUS Porous solids INDEX ON/OFF Index calculation on/off REFERENCE 8.000 Reference density for the index calculation in g/ccm (only used if REFERENCE <>...
■ 14 Density ("DENSITY") 14.5 Density of a liquid "MODE LIQUID" (with density kit 350-8515) This method is used to determine the density of a liquid. A glass vessel is used for this with a volume of 10 or 100 cm Display Step 0,000...
14 Density ("DENSITY") ■ 14.6 Density of a solid "MODE SOLID IN AIR" This method is used to determine the density of a solid. The double beaker is required for this. The thermometer is used to monitor the temperature of the test liquid in the container. Display Step 0,000...
■ 14 Density ("DENSITY") 14.7 Density of a porous solid "MODE SOLID POROUS" This method is used to determine the density of a porous solid. The double beaker is required for this. The thermometer is used to monitor the temperature of the test liquid in the container. Display Step 0,000...
14 Density ("DENSITY") ■ 14.8 Density of a solid "MODE SOLID ON THE BOTTOM" This method is used to determine the density of a solid; however, no auxiliary apparatus is required. The thermometer is used to monitor the temperature of the test liquid. Display Step 0,000...
■ 15 Differential weighing ("DIFF.-WEIGHT") Differential weighing ("DIFF.-WEIGHT") 15.1 Explanatory notes on differential weighing In the application differential weighing, samples are investigated for weight changes, the results logged in a report and collated in statistics. In differential weighing the original weight of a sample is determined at the start of a measurement. Following the treatment of the sample, e.g.
15 Differential weighing ("DIFF.-WEIGHT") ■ UNIT WEIGHT LOSS in the current unit of weight LOSS % RESIDUAL % ATRO 1 ATRO 2 CALCULATED DECIMALS FACTOR n.nnn e+n only if the "CALCULATED" unit has been selected MODE F * DIFF. only if the "CALCULATED" unit has been F / DIFF.
15 Differential weighing ("DIFF.-WEIGHT") ■ 15.4.1 Mode, measurement recording There are four different options available for semiautomatic measurement recording. Key: T: Tare I: Original weight R1-R3: Residual weight 1-3 Single The tare weight, original weight and residual weight are recorded in consecutive order for each sample.
■ 15 Differential weighing ("DIFF.-WEIGHT") If a reading has not been taken, the user cannot move on to the next one yet. Navigation in single mode The navigation for residual weights 2 and 3 must be done manually. The user gets to the next sample by pressing the «...
15 Differential weighing ("DIFF.-WEIGHT") ■ Display Step 15,85000 «E» Accept tare weight 1. 1-1 t + 15,85000 9 15,85000 Fill container 1 with original weight 1 and place it on. 1-1 I + 0,00000 9 56,10000 «E» Accept original weight 1. 1-1 i + 40,25000 9 As group 1 mode is set, the program automatically switches to the tare measurement for sample 2.
■ 15 Differential weighing ("DIFF.-WEIGHT") 15.4.4 Deleting samples, groups Press the «DEL» function key to go to the Delete submenu in the main menu. - - - - - - - - Delete submenu Configuration of the function keys: Functions Delete a single sample.
15 Differential weighing ("DIFF.-WEIGHT") ■ current unit. If the result is displayed, the differential weighing report is printed. Date and time, if they are switched on. (Can be set in the Configuration menu) Date 10.04.2004 Time 10:05:30 Balance ID, if it is switched on. (Can be set in the Configuration menu) Name : XR 125 SN Software...
■ 16 Minimum original weight ("MIN.-WEIGHT") Minimum original weight ("MIN.-WEIGHT") 16.1 Explanatory notes on the minimum sample weight application, MSW MSW application is a minimum original weight solution which enables you to fulfill QM guidelines, such as GLP, GMP or USP. 16.1.1 Minimum original weight and quality management Only very small quantities are used in many applications and, thus, only a small part of the balance's weighing capacity is used.
16 Minimum original weight ("MIN.-WEIGHT") ■ 16.3 Configuring the minimum original weight • SETUP APPLICATION MIN.-WEIGHT INFO LINE ON/OFF Display the info line permanently ZERO KEY ON/OFF Facilitate resetting SET PRINT FORMAT MIN. WEIGHT ON/OFF VIEW / SET PARAMETERS FLOATINGDISPLAY 0.16 (CODE - - - - ) STABILITY...
■ 16 Minimum original weight ("MIN.-WEIGHT") if there is a decisive change in the weighing criteria (ambient and application conditions). Once the date has elapsed, a warning briefly appears in the balance display when the MSW application is started. "TEST-PARA." You can enter a text here which describes the parameters which you have used for determining the minimum original weight (extension factor, uncertainty).
16 Minimum original weight ("MIN.-WEIGHT") ■ • "SET PRINT FORMAT" – "MIN. WEIGHT ON" Min.Wgt. 0.07500 g Minimum original weight has not been reached yet Actual : *+ 0.01532 g Min.Wgt. 0.07500 g Minimum original weight has been reached Actual : + 0.07874 g 16.4.2 Displaying additional information with the function key MIN...
Page 56
■ 16 Minimum original weight ("MIN.-WEIGHT") • Extension factor (k) (generally 2 or 3): The extension factor determines the likelihood of occurrence. The factor is defined by the user or is defaulted. The minimum original weight is calculated as follows: ⋅...
■ 17 Dynamic weighing ("DYN.-WEIGHT") Dynamic weighing ("DYN.-WEIGHT") 17.1 Introduction to the dynamic weighing application The "DYN. WEIGHT" workflow supports accurate weighing on an unstable surface, e.g. on board ships. During the period of measurement defined by the user in the setup menu, the balance measures continuously, calculates the mean of the saved values at the end of the period of measurement and indicates the average measurement value which this yields.
18 Pipette calibration ("PIPETTE") ■ Pipette calibration ("PIPETTE") 18.1 Introduction to pipette calibration Quality management systems require volumetric measuring equipment to be checked regularly. This application allows you to check pipettes of any make by means of a gravimetric method. You require a thermometer and a barometer in order to conduct this.
■ 18 Pipette calibration ("PIPETTE") "SAMPLES" Enter the number of random sample measurements per volume (setting range: 2-50). "EVAPORATION MEAS." – ON: The evaporation over the cycle time is automatically determined before the pipetting process starts, after the tare vessel is placed on, and the pipetting volumes are thus corrected accordingly. –...
18 Pipette calibration ("PIPETTE") ■ 18.4.1 Conducting the pipette test Press «esc» to stop the procedure at any time. Display Step - - - - - - - - «E» If necessary, change the (starting) temperature Temperature 22.0C - - - - - - - - «E»...
Page 62
■ 18 Pipette calibration ("PIPETTE") Display Step 0,00000 « » « » Pipette test failed. FAILED Press the «PRINT» key to print out the report on the pipette test or to send it to the PC. Press «INF» (« ») to display the results on the balance display as well: Display Step 0,00000...
■ 19 Air buoyancy correction ("BEST") Air buoyancy correction ("BEST") 19.1 Introduction to air buoyancy correction The "BEST" (Buoyancy Error Suppression Technology) application can be used to correct errors which arise as a result of air buoyancy. Balances are adjusted with steel weights with a density of 8 g/cm to prevent errors occurring at this density.
19 Air buoyancy correction ("BEST") ■ 19.4.1 Weighing weights Once you have accepted the current air density by pressing «OK», the balance can be used as in normal weighing mode, i.e. taring with «T» etc., for example. For marking purposes, the weight display contains the weight reading, which has been offset against the air buoyancy correction;...
■ 20 User profiles (MUM, Multiuser Memory) User profiles (MUM, Multiuser Memory) 10 different user profiles can be saved. They can be protected against changes by means of a personal 4- digit password. A profile consists of the configuration and application settings. Anyone who does not wish to create a personal user profile can work with the balance as a “guest“.
20 User profiles (MUM, Multiuser Memory) ■ The user is defined. Press «esc» to exit the menu. If there is a password set, it must be entered before making changes in the configuration and application menus. NOTE Make a note of your personal password. If a user loses his password, he can be enabled again using the password 7 9 1 4.
■ 21 Data transfer Data transfer For data-transfers to peripheral devices, the balance is equipped with an RS232/V24-interface. Before the data-transfer, the RS232 interface must be matched with the one in the peripheral device in the balance configuration menu (see chapter 10.3.6 "Interface functions"). •...
21 Data transfer ■ • Standard, duplex connection with additional hardware handshake in the peripheral device Balance DB 9 female D25 / D9 Peripheral device RS 232 out RS 232 in RS 232 in RS 232 out 21.2 Remote control-commands Command Function ACKn...
■ 21 Data transfer NOTE Each remote control-command must terminate with «CR» «LF». The commands are acknowledged if re- quired. 21.2.1 Examples for the remote control Input Description of the function executed D – – – – – Five dashes are displayed DTEST123 tESt123 is displayed The display is dark...
22 Maintenance and servicing ■ Maintenance and servicing 22.1 Calibration The calibration of the balance is fixed in the Configuration menu (see chapter 9.7 "Calibration of the balance" and see chapter 10.3.4 "Calibration functions"). Possible types of calibration, depending on the model of balance: •...
■ 22 Maintenance and servicing Then proceed as follows: Display Step +0,0000 The balance is in Weighing mode. +0,0000 «T» Press until „CALIBRATION“ apears. CALIBRATION - - 0000 The balance carries out a Zero measurement „0000 g“ is shown flashing). - - 112 After the zero measurement the display flashes with the pre- viously entered calibration weight.
22 Maintenance and servicing ■ Display Step +0,0000 «T» Press until „REPEATABILITY TEST“ apears. Repeatability test - - 000 The internal weight is applied and measured. 10 measurements are performed. measurement 1 EXIt - - 000 « » If required, the measuring program can be exited. measurement 3 EXIt +0,0001...
■ 22 Maintenance and servicing 22.5 Error messages The balance shows a description of the fault in the info-line. NOTE If an error occurs without a description of the error in the info-line, the Customer Service must be called. 22.5.1 Notes on correcting faults The following table shows faults and their possible causes.
23 Menu trees ■ Menu trees 23.1 Configuration menu tree • UNIT-1 UNIT-1 • SET DATA PRINT AUTOSTART ON/OFF MODE UNSTABLE STABLE LOADCHANGE CONTINUOUS TIMEBASE TIMEBASE SET PRINTFORMAT DATE AND TIME ON/OFF BALANCE-ID ON/OFF PRODUCT-ID ON/OFF GROSS AND TARE ON/OFF UNITS ON/OFF USER...
Page 76
■ 23 Menu trees • SET INTERFACE BAUDRATE 1200 2400 4800 9600 19200 38400 57600 PARITY 7-EVEN-1STOP 7-ODD-1STOP 7-NO-2STOP 8-NO-1STOP 8-EVEN-1STOP 8-ODD-1STOP HANDSHAKE XON-XOFF HARDWARE PC DIRECT MODE OFF/ON • SET DATE AND TIME TIME [HH.MM.SS] DATE [DD.MM.YY] FORMAT STANDARD/US •...
23 Menu trees ■ 23.2 Application menu tree • SET APP. UNITS COUNT PERCENT CALCULATOR Other applications available, siehe Kap. 11 "Application". PAPER NET-TOTAL ANIMAL etc. • SETUP APPLICATION Division depends on the current application, siehe Kap. 11 "Application". • SET STATISTICS MODE STATISTICS RECORDER...