Keithley 2308 Quick Start Manual page 27

Portable device battery/charger simulator
Hide thumbs Also See for 2308:
Table of Contents

Advertisement

20
Front-panel operation
Programming examples: pulse current measurements
The following command sequence will return the average of 10 peak pulse current
measurements:
Battery channel (#1)
SENS:RANG 5
VOLT 15
CURR 0.75
OUTP ON
SENS:PCUR:SYNC ON
SENS:PCUR:AVER 10
SENS:PCUR:SYNC:TLEV 0.1
SENS:PCUR:TIME:AUTO
SENS:PCUR:SYNC:DEL 50e-3
SENS:FUNC "PCUR"
SENS:PCUR:MODE HIGH
READ?
Charger channel (#2)
SENS2:RANG 5
SOUR2:VOLT 15
SOUR2:CURR 0.75
OUTP2 ON
SENS2:PCUR:SYNC ON
SENS2:PCUR:AVER 10
SENS2:PCUR:SYNC:TLEV 0.1
SENS2:PCUR:TIME:AUTO
SENS2:PCUR:SYNC:DEL 50e-3
SENS2:FUNC "PCUR"
SENS2:PCUR:MODE HIGH
READ2
' Select 5 A range.
' Set output voltage to 15 V.
' Set current limit to 75 0mA.
' Turn output on.
' Enable trigger synchronization.
' Set average count to 10.
' Set trigger level to 100 mA for 5 A trigger
level range.
' Set integration times automatically.
' Set trigger delay to 50 msec.
' Select pulse current function.
' Configure to measure peak pulse.
' Trigger 10 measurement conversions and
return the average of those 10 conversions.
The average of the 10 conversions is
displayed on the front panel. Each of the 10
conversions syncs to the rising edge.
' Select 5 A range.
' Set output voltage to 15 V.
' Set current limit to 750 mA.
' Turn output on.
' Enable trigger synchronization.
' Set average count to 10.
' Set trigger level to 100 mA.
' Set integration times automatically.
' Set trigger delay to 50 msec.
' Select pulse current function.
' Configure to measure peak pulse (trigger
on).
' Trigger 10 measurement conversions and
return the average of those 10 conversions.
The average of the 10 conversions is
displayed on the front panel. Each of the ten
conversions syncs to the rising edge.