Keithley 2306NV User Manual
Keithley 2306NV User Manual

Keithley 2306NV User Manual

Battery/charger simulator

Advertisement

Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
(440) 248-0400
Fax: (440) 248-6168
www.keithley.com
NOTE
This information is to be used in conjunction with the Model 2306 Instruction manual. The
Instruction manual is on the Product Information CD-ROM that ships with the Model 2306NV.
Make sure to thoroughly understand the operation of the Model 2306NV before operating or con-
necting the unit.

Overview

This User's Guide contains operational differences between the Model 2306NV and the standard Model 2306. The model num-
ber displayed during the power up sequence is 2306NV instead of 2306. The REVISION NUMBER under MAIN MENU dis-
plays 'NV' between the two revision numbers as shown below:
REVISION NUMBER
B13
NV B01
Additionally, sending the identification query ' *IDN? ' returns the 2306NV model number as shown below:
KEITHLEY INSTRUMENTS INC., MODEL 2306NV, xxxxxxx, yyyyy/zzzzz
xxxxxxx is the serial number.
Where:
yyyyy is the firmware revision level of the digital board ROM.
zzzzz is the firmware revision level display board ROM.

2306NV operation

Essentially, the Model 2306NV functions the same as the Model 2306 with one exception: the Model 2306 can perform DVM
measurements while the 2306NV (no voltmeter) cannot. With this in mind, note the following differences:
• The DVM menu item (DVM INPUT) has been removed. Also, the quick disconnect connectors contain non-functional
DVM + and DVM - terminals.
• The commands contained in Table 1 have changed. Sending these unsupported commands will cause errors.
Table 1
2306NV Command differences
Type
Independent voltage
measurements (DVM)
Signal Oriented
Common Commands
Model 2306NV Battery/Charger Simulator
Command
:SENSe[1]:FUNCtion <name>
:SENSe2:FUNCtion <name>
:MEASure[1][:<function>]?
:MEASure[1]:ARRay[:<function>]?
:MEASure2[:<function>]?
:MEASure2:ARRay[:<function>]?
*IDN?
Change
Command <name> of 'DVM' and 'DVMeter'
removed. Also, the portion of associated commands
such as :NPLCycles <n> or :AVERage <NRf> with
respect to DVMeter, have also been removed.
Command <function> 'DVM' and 'DVMeter'
removed.
This command returns 2306NV for model type.
User's Guide
PA-855 Rev. B / 1-05

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 2306NV and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keithley 2306NV

  • Page 1 firmware revision level display board ROM. 2306NV operation Essentially, the Model 2306NV functions the same as the Model 2306 with one exception: the Model 2306 can perform DVM measurements while the 2306NV (no voltmeter) cannot. With this in mind, note the following differences: •...
  • Page 2 Calibration When performing calibration on a Model 2306NV, use the procedure as contained in the Model 2306 Instruction Manual. Take care to include the call for calibrating the DVM (this call must be made or an error will result). Verification...

Table of Contents