Variable output impedance control on Battery Channel (#1)
Variable output impedance control on battery
channel
The Model 2308 battery channel (#1) has a variable output
impedance control that can be used to simulate the impedance
of a battery pack.
To manually set the output impedance:
1. Press the MENU key.
2. Press the
3. Press the ENTER key and use the
4. Press the ENTER key to program the value.
5. Press the MENU key to exit from the menu options.
To program the output impedance from a remote interface:
Send the
NOTE
or
keys to scroll through the menu items until OUTPUT
▲
▼
IMPEDANCE is displayed.
impedance between 0.00 Ω and 1.00 Ω in 0.01 Ω steps.
OUTput:IMPedance<NRF>
The output impedance can be changed while the output is on.
or
keys to set an output
▲
▼
,
,
,
command over the GPIB bus.