Nikon eclipse fn1 Instructions Manual page 36

Table of Contents

Advertisement

(2)
Attaching the IR Polarizer
1
Separate the upper part (band pass filter and 1/4
lambda plate) and the lower part of the IR
polarizer.
2
Insert the adjustment tool for the 1/4 lambda plate
orientation, which is provided with the polarizer
turret, into the lower part of the IR polarizer.
In doing so, adjust the notch position on the
adjustment tool for the 1/4 lambda plate and the
protrusion position on the lower part of the IR
polarizer.
3
Rotate the knurled ring on the right side of the polarizer turret to adjust the index (a white
point) to the center line of the scale. Insert the IR polarizer assembled in Step 2 into the upper
turret of the polarizer turret. The notch of the IR polarizer must be oriented in the way as
shown in the figure.
Notch
4
Attach the polarizer turret into the microscope body.
5
Rotate the knurled ring on the right side of the polarizer turret and search the position where
the view becomes darkest. After the adjustment, tighten the clamp screw.
Chapter 3
Operation of Each Part
11
Polarizer Turret
34
Upper part
(Band pass filter and 1/4 lambda plate)
Lower part
Adjustment
tool for the 1/4
lambda plate
orientation
Lower part of the IR polarizer
Clamp screw
Knurled ring
Notch
Protrusion
Align the white
point and the
center line.