MAN 01/04 CRONO FE 03/11
DISPlaYInG THE nUMbER OF COMPlETE
•
The pump has to be in OFF to be able to display
the number of complete infusions performed by
the pump.
•
Depress the
number of complete infusions will be displayed.
lOW baTTERY alERT
The LOW BATTERY symbol will be continuously dis-
played when the battery level is low.
The battery should be changed as soon as the
ongoing infusion is completed.
If the battery is completely discharged, BATTERY
DISCHARGED, the symbol batt will be displayed and
the pump will emit a short acoustic signal. The pump
will stop and the infusion cannot be completed
without changing battery.
•
Avoid that the pump stops during an ongoing infu-
sion by changing the battery as soon as the LOW
BATTERY symbol appears and the ongoing infu-
sion is completed.
•
Do not replace the battery:
-
during an ongoing infusion
-
when the infusion set is connected to the patient
InFUSIOnS
button for about 7 seconds and the
CaUTIOnS