PROGRaMMInG OF F-bOlUS DOSE VOlUME

The F-bolus dose volume can be programmed from
5 µl to 1000 µl in steps of 5 µl.
The F-bolus dose is accessed by the second depres-
sing of the
sequence (see page 25).
Programming procedure:
1.
When the
time, the letter "d" is displayed and the presently
programmed F-bolus dose is flashing.
2.
By depressing the
se; by depressing the
decrease. Any change is followed by an acoustic
signal.
3.
If no button is depressed within 5 seconds in the
programming phase, the value stops flashing and
OFF appears on the screen.
4.
Depressing the
ge appears (the F-bolus dose is still flashing)
moves the programming sequence to the next
option:
interval time between F-bolus doses.
-
Programmed F-bolus dose is automatically stored.
-
You can accelerate (scrolling) the display of new
value by depressing the
continuously (in steps of 10 µl)
-
If iIt is not necessary to re-program the present
value, just depress the
flashes.
button in the pump programming
button is depressed for a second
button the value will increa-
button the value will
button before the OFF messa-
nOTE
or the
button while the value
button
27
10
20
10
2 0
10
2 0
10
2 0