Starting The Pump; Priming - Cane Crono User Manual

Ambulatory drug delivery pump
Table of Contents

Advertisement

MAN 01/04 CRONO FE 03/11
Depress the
there will be a short acoustic signal and the display
will show either:
-
Pr (priming function) provided that the pusher is in
the start position for a new, full or partial volume,
infusion
or
-
remaining infusion time or remaining time to
the next F-bolus.
The display shows the remaining delivery time to the
end of the infusion or the remaining time to the next
F-bolus in minutes.
The remaining time to the end of the infusion is dis-
played in days until less than one day remains, in
hours until less than one hour remains, then in minu-
tes.
The priming function is used for filling the infusion set
with drug from the syringe. A total of 3 ml, in steps of
0.5 ml, can be primed using the priming function.
Priming can only take place when the pusher is in the
starting position for a new, full or partial volume, infu-
sion.

STaRTInG THE PUMP

button when the pump is in OFF and
nOTE

PRIMInG

31
10
20
10
2 0