Pump In On (Running); F-Bolus Dose - Cane Crono User Manual

Ambulatory drug delivery pump
Table of Contents

Advertisement

MAN 01/04 CRONO FE 03/11
The display will show Pr and alternative a,
options can be selected again.
There is no time limit to finish the priming procedure.
The priming volume is not included in the administra-
tion volume.

PUMP In On (RUnnInG)

The display shows the remaining delivery time to the
end of the infusion or the remaining time to the next
F-bolus in minutes.
The remaining time to the end of the infusion is dis-
played in days until less than one day remains, in
hours until less than one hour remains, then in minu-
tes.
The F-bolus dose is automatically administered as
programmed. Both the volume and the interval time
is pre-programmed and the patient does not have to
perform any operational functions.
During the administration of an F-bolus dose the
administered volume in µl will be displayed.
The duration of a 5 µl F-bolus is about 0.5 seconds,
the duration of 1000 µl F-bolus is about 1'35 ".
During the administration of an F-bolus dose the
visualisation function are not active.
The pump must not be switched off during the admi-
nistration of an F-bolus; if it is switched off and afte-
wards switched on the F-bolus will not be finished.
nOTE

F-bOlUS DOSE

b
or
c
33
10
2 0
10
2 0
10
20
10
2 0