PROGRaMMInG OF PaRTIal VOlUME

Partial volume is used when less than 10 or 20 ml is
drawn up in the syringe for a specific treatment or for
a specific patient.
Partial volume can be programmed between 1 ml to
20 ml with steps of 1ml.
Partial volume is accessed by the fourth depressing
of the
button in the pump programming sequen-
ce (see page 25). Please observe that partial volume
is only available at the beginning of a new partial
infusion.
Programming procedure:
1.
When the
the presently programmed volume in ml is flas-
hing.
2.
By depressing the
se; by depressing the
decrease. Any change is followed by an acou-
stic signal.
3.
If no button is depressed within 5 seconds in the
programming phase, the selected volume in ml
stops flashing and P,cc appears on the screen.
4.
The pusher will automatically position itself for the
button while the value
button is depressed for a fourth time,
button the value will increa-
button the value will