InTERVal TIME bETWEEn F-bOlUS DOSES

Interval time between F-bolus doses can be program-
med from 5 minutes to 300 minutes in steps of 5 minu-
tes.
Interval time between F-bolus doses is accessed by
the third depressing of the
gramming sequence (see page 25)
Programming procedure:
1.
When the
button is depressed for a third time,
a "t" is displayed and the presently programmed
interval time between F-boluses is flashing.
2.
By depressing the
se; by depressing the
decrease. Any change is followed by an acoustic
signal.
3.
If no button is depressed within 5 seconds in the
programming phase, the interval time between A-
Boluses stops flashing and OFF appears on the
screen.
4.
Depressing the
ge appears (the interval time between F-boluses is
still flashing) moves the programming sequence to
the next option: partial volume.
-
Programmed interval time between F-boluses is
automatically stored.
-
You can accelerate (scrolling) the display of new
values by depressing the
28
MAN 01/04 CRONO FE 03/11
button in the pump pro-
button the value will increa-
button the value will
button before the OFF messa-
nOTE
or the
button conti-