Electromagnetic Immunity And Emissions Declaration; Immunity Standard; Emissions Summary - Welch Allyn SureTemp Plus 690 Service Manual

Welch allyn suretemp plus service manual
Hide thumbs Also See for SureTemp Plus 690:
Table of Contents

Advertisement

Electromagnetic Immunity and Emissions Declaration

Immunity Standard

IEC Basic Standard
Electrostatic discharge IEC 61000-4-2
RF EMF Immunity IEC 61000-4-3
Magnetic field IEC 61000-4-8

Emissions Summary

The SureTemp
electromagnetic environment specified below. The customer or the user of the model 690 and
692 should assure that it is used in such an environment.
Guidance and Manufacturer's declaration - electromagnetic emissions
Emissions test
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
and flicker emissions
IEC 61000-3-3
SureTemp® Plus Service Manual
®
Plus model 690 and 692 thermometers are intended for use in the
Compliance
Group 1
Class B
Complies
Complies
IEC 60601-1-2 2001
± 6kV
c
± 8kV
c
3/10 V/m
80 MHz-2.5GHz
3A/m
150kHz-80MHz
Electromagnetic
environment-guidance
®
The SureTemp
Plus model
690 and 692 Thermometers
are suitable for use in all
establishments, including
establishments that are
connected to the public
low-voltage power supply
network.
1

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SureTemp Plus 690 and is the answer not in the manual?

This manual is also suitable for:

Suretemp plus 692

Table of Contents

Save PDF