Australia “C-Tick” Symbols and Description Consult directions for use (DFU). A copy of the DFU is available on this website. A printed copy of the DFU can be ordered from Welch Allyn for delivery within 7 calendar days welchallyn.com Type BF Applied Part Fragile +55°C...
Failure to use a probe cover can cause patient discomfort from a heated probe, patient cross-contamination, and inaccurate temperature readings. WARNING Do not take a patient’s temperature without using a Welch Allyn disposable probe cover. Doing so can cause patient discomfort, patient cross contamination, and erroneous temperature readings.
Page 7
Caution Welch Allyn recommends that batteries are removed for long term storage since alkaline batteries can deteriorate and leak over extended periods of time, and possibly cause damage to the thermometer.
Parts and Features You can find the serial number of your device on a label affixed to the rear housing inside the unit’s probe cover storage compartment. Please note the parts of your Welch Allyn SureTemp Plus thermometer as shown below:...
“snaps” into position and is completely seated to the thermometer. Place the probe into the well. 3. Insert the box of Welch Allyn disposable probe covers into the probe cover storage holder. To open the box of probe covers, remove and discard the tear-away corner. You...
Page 10
Welch Allyn SureTemp Plus Model 690V MOUNTING THE UNIT (with optional wall mount) Mount the wall-mount bracket in a location such as an interior wall or wall divider so that the thermometer remains at ambient room temperature. Mounting the unit on an...
The device is not working properly. Contact Welch Allyn’s Customer Service Department for assistance. Broken Probe The probe is broken. Replace probe or contact Welch Allyn’s Customer Service Department for assistance. Loss of Tissue Contact The probe has lost contact with the patient’s tissue. The icon will disappear once proper contact is achieved.
Page 12
Welch Allyn SureTemp Plus Model 690V Ambient Temperature Exceeds Upper Measurement Range Limit of 104.0 °F (40.0 °C) Ambient Temperature Is Lower than Measurement Range Limit of 50.0 °F (10.0 °C) Low Battery A single flashing segment appears on the battery symbol.
When used correctly, the SureTemp Plus thermometer accurately measures rectal temperature in approximately 10–13 seconds. TAKING A RECTAL TEMPERATURE WARNING Do not take a patient’s temperature without using a Welch Allyn disposable probe cover. Doing so can cause patient discomfort, patient cross contamination, and erroneous temperature readings.
3. Load a probe cover by inserting the probe into a probe cover and pressing the probe handle down firmly. The probe handle will move slightly to engage the probe cover. WARNING Use only Welch Allyn probe covers. The use of other manufacturer’s probe covers or no probe cover may produce temperature measurement errors and/or inaccuracy.
Directions for use 10. Replace the probe in the probe well to clear the display and reset the thermometer to Normal Mode. The thermometer will not beep to indicate a final temperature. WARNING The unit will automatically enter Monitor Mode if the probe is withdrawn from the probe well and is not replaced within 60 seconds of inactivity.
Welch Allyn SureTemp Plus Model 690V CLEANING AND DISINFECTING THE REMOVABLE PROBE WELL • Remove the probe well from the unit. Unplug the latching probe connector to prevent the device from consuming battery power while you are cleaning the probe well.
Directions for use Performance and Technical Specifications 50.0 F to 104.0 F (10.0 C to 40.0C) Ambient Operating Temperature Range Operating Altitude -557 ft. to +16,000 ft / -170m to +4877m 80.0 F to 110.0 F (26.7 C to 43.3 C) Patient Temperature Range Temperature Predict Time* Rectal:...
Electromagnetic Emissions The Welch Allyn SureTemp Plus model 690V is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch Allyn thermometer should assure that it is used in such an environment. Emissions Test...
Page 19
Electromagnetic Immunity The Welch Allyn SureTemp Plus model 690V is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn thermometer should assure that it is used in such an environment. Immunity Test...
Page 20
Electromagnetic Immunity The Welch Allyn SureTemp Plus models 690V is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn thermometer should assure that it is used in such an environment. Immunity Test...
Page 21
Welch Allyn SureTemp Plus model 690V The Welch Allyn SureTemp Plus model 690V is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Welch Allyn thermometer can help prevent...
Before returning a product for repair, you must first obtain authorization from Welch Allyn. If you are advised to return a product to Welch Allyn for service or repair, you should schedule the repair with the Welch Allyn Service Center nearest you.
Model 690 from the date of first shipment from Welch Allyn. This warranty shall be fulfilled by Welch Allyn or its authorized representative repairing or replacing at Welch Allyn’s discretion, any such defect, free of charge for parts and labor.