6. Press the container with the
control solution so that a tiny
droplet forms on the dosing
tip. Place the drop on a
surface that you can reach
easily with the test strip (e.g.
the hollow of the screw cap).
7. Place the blood sample area
of the test strip (capillary) on
the drop. Make sure that the
test strip's capillary is
completely filled.
8. The device starts a reading.
After about 5 seconds, the
test result of the reading with
the control solution appears.
9. Before removing the test
strip, see whether the test
result is within the range
specified on the test strip
container.
Sample values:
Normal: 104 - 140 mg/dL