Reset; Resetting Calibration Values - wtw pH 3110 Operating Manual

Ph meter
Hide thumbs Also See for pH 3110:
Table of Contents

Advertisement

Operation
Calibration values in the
default condition
Resetting calibration
values
34
4.6

Reset

4.6.1 Resetting calibration values

This function resets the calibration values to the default condition. All
other meter settings are retained.
Zero point
Slope
Note
The measuring system is not calibrated after a reset. Before
measuring, recalibrate the meter.
1
Press <On/Off
calibration data.
Init.C is displayed.
pH
I
n
i
t
n
2
Press <▲><▼> to display no or YES.
YES: Reset the calibration values.
no: Retain the calibration values.
3
Confirm with <ENTER>.
The menu is finished.
The meter switches to the measuring mode.
0 mV (pH 7.000)
-59.16 mV/pH (100 %)
_
> to open the menu for the reset of the
C
o
pH 3110
ba75791e02
11/2008