wtw pH 3110 Operating Manual page 25

Ph meter
Hide thumbs Also See for pH 3110:
Table of Contents

Advertisement

pH 3110
ba75791e02
11/2008
A
S
2
4
ConCal
2
If necessary, set the temperature of the first buffer solution with
<▲><▼>.
3
Immerse the pH combination electrode in the first buffer
solution (pH 7.0 ± 0.5 for two-point calibration).
4
Press the <ENTER> key.
The measured pH value appears on the display.
5
When the measured value is stable,
using <▲><▼>, set the measured value to the nominal pH
value of the buffer solution (at the current temperature).
6
Press the <ENTER> key. SLO appears on the display.
Note
Here you can cancel the calibration procedure with <M>. This
corresponds to a single-point calibration. The value of the zero point
(ASY) is displayed. The sensor symbol shows the evaluation of the
single-point calibration. Pressing <ENTER> displays the value of the
slope (SLO).
s
L
2
4
ConCal
Y
8
TP
°C
O
8
TP
°C
Operation
25