pH 3110
Continuing with two-
point calibration
Continuing with three-
point calibration
ba75791e02
11/2008
Note
Here you can cancel the calibration procedure with <M>. This
corresponds to a single-point calibration. The value of the zero point
(ASY) is displayed. Pressing <ENTER> displays the value of the slope
(SLO).
5
Thoroughly rinse the combination electrode with distilled water.
6
If necessary, set the temperature of the second buffer solution
with <▲><▼>.
7
Immerse the combination electrode in the second buffer
solution.
8
Press the <ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the nominal value of the buffer
(setting: see section 4.5.2) is displayed. As soon as a stable
value is recognized, Ct3 or Cd3 appears.
pH
C
T
3
2
8
4
AutoCal TEC
Note
Here you can cancel the calibration procedure with <M>. This
corresponds to a two-point calibration. The value of the zero point
(ASY) is displayed. Pressing <ENTER> displays the value of the slope
(SLO).
9
Thoroughly rinse the combination electrode with distilled water.
10
If necessary, set the temperature of the third buffer solution
with <▲><▼>.
11
Immerse the combination electrode in the third buffer solution.
° C
TP
AutoCal DIN
Operation
pH
C
3
D
2
8
4
° C
TP
23