Annual Product Inspection; Reissuing Of Kidwalk; Product Servicing - Prime Engineering KidWalk Owner's Manual

Dynamic mobility system
Table of Contents

Advertisement

Annual Product Inspection

(Therapist, Technician, Dealer)
It is recommended that each KidWalk should be
subject to a detailed inspection at a minimum of
least once a year and every time the product is reis-
sued for use. This inspection should be carried out
by a technically competent person who has been
trained in the use of the product and should include
the following checks as a minimum requirement.
1. Verify that the seat height adjustment mechanism
is locking securely.
2. Check the mast secures to the base frame and the
pin moves freely into the locked position.
3. Verify that all knobs, nuts, bolts, plastic buckles
and Velcro are in place and have any missing items
replaced.
4. Lift the base to check each wheel individually.
Make sure they are moving freely and remove any
dirt from the rubber wheels.
5. Check that the brakes lock the wheels fully.
5. Visually check the structure of the product paying
attention to the frame ensuring there are no signs of
fatigue or cracking around the welds or on the
frame.
6. Make sure that all nuts and bolts are tight and
secure on all of the components.
6. Maintain a written record of all product inspec-
tions. If there is any doubt of the continued safe
usability of your KidWalk or if any parts should fail,
cease using the product immediately and contact our
customer services department or your local dealer as
soon as possible.
7. Check that the mast tilt adjustment mechanism is
working correctly.

Product Servicing

Servicing of any KidWalk should only be done by technically competent persons who have
been trained in the use of the product.
KidWalk
Owner's Manual
Reissuing KidWalk
Most KidWalk's are ordered and sized to meet the
needs of a specific individual. Before reissuing a
KidWalk the therapist prescribing the product
should have carried out an equipment compatibil-
ity check for the new client and has ensured that
the product being re-issued contains no modifica-
tions or special attachments. The actual compo-
nents of the current unit should also be verified as
to the correct fit for the new client as well.
A detailed technical inspection should be carried
on the product prior to re-issuing. This should be
carried out by a technically competent person who
has been trained in the use and inspection of the
product. Please refer to the Annual Product
Inspection section for the required checks.
Ensure the product has been cleaned thoroughly in
accordance with the Cleaning and Care section of
this manual.
Ensure a copy of the Owner's Manual is supplied
with the product. If one is not with the unit you
can request one from Info@primeengineering.com
or your supplying dealer.
A written record should be maintained of any and
all product inspections and changes done during
any reissue of the KidWalk.
If in any doubt to the continued safe use of your
KidWalk or if any parts should fail, please cease
using it immediately and contact our customer
service department or your local dealer as soon as
possible.
17

Advertisement

Table of Contents
loading

Table of Contents